1.3Model 2002 Control Unit
The Model 2002 control unit consists of the power supply, Pirani resistance bridge control, signal processing and display circuitry. The user interface consists of a flat panel display with smooth operat- ing tactile switches and large green LED digits (0.56"/14.2mm). The pressure is displayed in scientific format using a 3 digit mantissa and a 1 digit (plus the sign) exponent. The pressure can be displayed in Torr, mbar or Pascal. The instrument comes standard with 2 alarm setpoints that have TTL level digital outputs. An optional dual relay board can be internally mounted to trip on the alarm setpoints to turn on/off various equipment such as pumps, valves, heaters, bakeout ovens and safety equipment. An optional
An optional serial interface board can be installed to provide RS232 or RS485 support. The Model 2002 control unit fits a standard DIN 43700 insert (1/4 DIN) and mounts into a 3.62 in. (92 mm) square hole. The case is constructed of NORYL rated
equipped with mounting brackets that are adjustable up to 3/16” (4.8mm).
1.4Specifications
| Measuring range | ||
|
| ||
| Accuracy (N , T=230C) | + 15% of reading | |
| 2 | + 1.5% of reading (3x10+1 to 1x10+3 Torr) | |
|
| ||
| Ambient temperature operating range | 00 to 500C | |
| Bakeout temperature (with transducer electronics removed) | 2500C | |
| Analog output | ||
| Process control | 2 TTL outputs (1 TTL auto zero input) | |
! | Digital readout | Four green LED - resolution is three digit | |
|
| (plus exponent) | |
| Equipment operating range | 120/240 V~, 0.25A, 50/60 Hz | |
|
| A 6 ft., | |
| Fuse | 250 V, 1/4 A. | |
| Transducer mounting | Any position without recalibration. | |
| Transducer internal volume | <1.5cc | |
| Wetted material | Au, Si | N , Si, Pt, PyrexTM, |
|
| 3 | 4 |
|
| KovarTM and 316 stainless steel | |
| Stability with voltage | Variation in reading undetectable | |
|
| as power fluctuates within power specification. | |
| Weight (display, cable, and sensor) | Net: 2.5 lbs., Shipping: 3.4 lbs. | |
| Transducer response time | Typically, less than 150 msec for a step | |
|
| change from | |
| Positive pressure | 150 psig, Recalibration may be required | |
| if unit is exposed to pressures greatly exceeding measuring range. |
page 6 Model 2002