Transmit Selected Units | U<CR> |
| Torr<CR> |
|
Transmit Software Version# | V<CR> | Hastings Instruments - Model 2002 | ||
|
| Version 1.60 - | ||
Parameter Modification Commands |
|
|
| |
Command Description | Format |
| Valid Range: |
|
Modify High Setpoint | H={m.dd}E{+e}<CR> | |||
Modify Low Setpoint | ||||
Modify Selected Gas # | G={d}<CR> |
| 0 To 4 | (Decimal) |
Modify Selected Units | U={u}<CR> |
| T, or M, {or P, if Available} | |
Modify Decimation Ratio | D={dddd}<CR> | 63 To 7936 | (Decimal) | |
Modify | *{aa}A={aa}<CR> | 1 To FF | (Hexadecimal) | |
Modify Turnaround Delay | *{aa}T={dd}<CR> | 0 To 255 | (Decimal) |
Notes:
The setpoints may also be entered as a decimal number, e.g. [H=760.99<CR>] will be same as entering [H=7.6099E+2<CR>] .
When inputting setpoint data, it should be entered in the same Units of Pressure as the presently selected Units of Measurement (i.e.Torr, mbar or Pascal).The data is only checked to be a valid number with a one digit exponent before being accepted.There are no limit checks on the data, the user is free to choose any value appropriate to his use of the instrument.
The Turnaround Delay and
If the
Caution: Since all units will respond to the ‘UNIVERSAL ADDRESS’, make sure that only the unit to be modified is connected to the
The value entered for the Turnaround Delay is used to modify an internal timer which normally runs at ~8 millisecond, e.g. [*{aa}T=10<CR>] will set the delay to ~80 milliseconds.
If the command syntax is not met or if the number is out or range, the
Calibration Adjustment Commands
Command Description | Format | Valid Range: |
page 32 Model 2002