Trace Oxygen Analyzer | Operation 4 | |
|
|
|
|
|
|
|
|
|
Use the DOWN/UP control to enter the upper value of the range (all ranges begin at 0 ppm). Repeat for each range you want to set. ENTER to accept the values and return to the Analysis Mode. (See note below.)
Note: The ranges must be increasing from low to high, for example, if range 1 is set as
Ranges, alarms, and spans are always set in ppm units (over the entire
After defining your analysis ranges, set Mode– to either AUTO or one of the fixed ranges, as described below.
4.10.2 Fixed Range Analysis
The autoranging mode of the instrument can be overridden, forcing the analyzer DC outputs to stay in a single predetermined range.
To switch from autoranging to fixed range analysis, ENTER the
RANGE function by selecting RANGE from the MAIN MENU.
Use the DOWN/UP | control to move AUTO between the arrows. |
Use the DOWN/UP | control to switch from AUTO to FX/LO, FX/MED, |
or FX/HI to set the instrument on the desired fixed range (low, medium, or high).
| or |
| or |
ESCAPE to
NOTE:When performing analysis on a fixed range, if the oxygen concentration rises above the upper limit, as established by the setup, for that particular range, the output saturates at 1 V dc (or 20 mA). However, the digital readout and the
Teledyne Analytical Instruments |