Operation Model 3110
36 Teledyne Analytical Instruments
To use this feature:
1. Use the UP/DOWN keys to navigate to the FILTER screen.
Then press ENTER to enter the setup screen.
2. Use the ENT/ON key to change the filter to the active mode.
The arrows will point left and right and the filter setting will
blink.
3. Use the UP/DOWN keys to select the desired setting.
4. Select the ENT/ON key to save the setting and to return to
the non-active mode on the filter screen.
3.2.16 LOG INTV Screen
This screen indicates the time interval between data samples taken by
the data logger. The interval can be set between 1 second (00m 01sec) to 60
minutes (60m 00sec) in increments of 1 second.
To change the interval between data samples:
1. Use the UP/DOWN keys to navigate to the LOG INTV
screen. Then press ENTER to enter the setup screen.
2. Use the UP /DOWN keys to change the interval value. The
value of the seconds will increase or decrease followed by
the minutes. Press ENTER to accept the displayed value and
return to the LOG INTV screen.
3.2.17 LOG RESET & START Screen
Once a log interval has been input, the LOG RESET & START
screen is used to start the data logger.
The data logger has a 3200 record capacity and each record uses 10
bits of data. The data set is retained until it is reset by using the START