4.2.1.SELECT
SELECT is a
Note: Ensure vacuum is below
4.2.2.ADJUST
ADJUST is a rotary encoder. Rotating ADJUST will select desired parameters within the different modes.
4.2.3.STATUS BAR (RED LED’s)
The status bar displays information about the condition of the
High and Low will illuminate when respective TTL set point is activated high (+5 VDC).
F1 and F2 illuminate when the respective filaments are turned on.
The error LED is located in the middle of the top row and is not labeled. It is illuminated if the filament is not turned on or the filament is turned on and indicated pressure is out of range.
4.2.4.MODE BAR (GREEN LED’s)
The Mode bar indicates in which mode the
A particular LED, when lit indicates that particular mode has been entered.
The Available modes are Emis (emission), Dgas, Sp (set point), K (K factor) and Fil filament).
Refer to Section 3.3 for detailed explanations of the modes.
4.3.Operating Modes
4.3.1.No Mode (Normal Mode):
NO MODE is the default (normal) operating mode of the
MODE the numeric display indicates BA tube out put (pressure measurement value). Dashes displayed on the numeric display indicate that the filament is not turned on, not operating properly, or that the readings are out of range. The filament will automatically turn off if indicated pressure is above 1x10
NOTE: Dashes displayed on the numeric display with the Filament 1 or Filament 2 LED’s illuminated, is an indication that the B/A tube is not operating properly.
4.3.2.Emis (emission) Mode:
The Emis mode is entered by scrolling to the Emis position by pressing the SELECT push button. The desired B/A tube emission current is then selected by rotating the ADJUST encoder to the desired emission range. Available settings are 0.01 mA, 0.1 mA, 1 mA or AUTO ranging. In most situations the optimum setting for the
IGE 300 | Page 11 of 21 |