4.2.1.SELECT

SELECT is a push-button. Pressing it will scroll through the different available operating modes. Rapidly “double clicking” SELECT will either turn the selected filament on (if unit is off) or off (if the unit is on)

Note: Ensure vacuum is below 1x10-2 Torr. The IGE-3000 has an automatic shutoff feature that will not allow filament turn on at pressures above 1x10-2 Torr. The shutoff feature will also automatically turn off the filament if it is on (below 1x10-2 Torr) and pressure rises above 1x10-2.

4.2.2.ADJUST

ADJUST is a rotary encoder. Rotating ADJUST will select desired parameters within the different modes.

4.2.3.STATUS BAR (RED LED’s)

The status bar displays information about the condition of the IGE-3000. A particular LED, when lit, indicates that particular condition has been met.

High and Low will illuminate when respective TTL set point is activated high (+5 VDC).

F1 and F2 illuminate when the respective filaments are turned on.

The error LED is located in the middle of the top row and is not labeled. It is illuminated if the filament is not turned on or the filament is turned on and indicated pressure is out of range.

4.2.4.MODE BAR (GREEN LED’s)

The Mode bar indicates in which mode the IGE-3000 is currently operating.

A particular LED, when lit indicates that particular mode has been entered.

The Available modes are Emis (emission), Dgas, Sp (set point), K (K factor) and Fil filament).

Refer to Section 3.3 for detailed explanations of the modes.

4.3.Operating Modes

4.3.1.No Mode (Normal Mode):

NO MODE is the default (normal) operating mode of the IGE-3000. During NO

MODE the numeric display indicates BA tube out put (pressure measurement value). Dashes displayed on the numeric display indicate that the filament is not turned on, not operating properly, or that the readings are out of range. The filament will automatically turn off if indicated pressure is above 1x10 -2. NO MODE is automatically entered when there are no modes selected or the current selected mode has timed out. Time out occurs if no operator action is initiated after 15 seconds.

NOTE: Dashes displayed on the numeric display with the Filament 1 or Filament 2 LED’s illuminated, is an indication that the B/A tube is not operating properly.

4.3.2.Emis (emission) Mode:

The Emis mode is entered by scrolling to the Emis position by pressing the SELECT push button. The desired B/A tube emission current is then selected by rotating the ADJUST encoder to the desired emission range. Available settings are 0.01 mA, 0.1 mA, 1 mA or AUTO ranging. In most situations the optimum setting for the IGE-3000 is AUTO ranging. In AUTO ranging the IGE-3000’s CPU automatically determines and sets the optimum emission current range as determined by the CPU and is based on the pressure measurement. When operating in other than AUTO ranging, the following settings are recommended: for the pressure ranges above 1x10-4Torr use 0.01 mA; for the pressure ranges 1x10-4to 1x10-5Torr use 0.1 mA and for the pressures below 1x10-5Torr use 1 mA.

IGE 300

Page 11 of 21

Page 11
Image 11
Teledyne IGE 3000 instruction manual Operating Modes