When entering more than one command in the same data string, they must be separated by a comma.

All command strings must be followed by the terminator character (carriage return <CR>, also known as ENTER).

When a lower case character is present in an example it represents an option.

Character Description Valid Inputs:

a RS-485 Address (hexadecimal 0-9, A-F) 01 - DF

m Most Significant Digit Of Mantissa 1 - 9 d Decimal Digit 0 - 9

e Exponent 0 - 5

u Unit Of Pressure T, M, or P

, Command Separator (comma) N/A

<CR> Command Terminator (carriage return) N/A

8.3.Interrogation Commands

Command Description

Format

Sample Response

Transmit K Factor

K<CR>

K Factor: 1.0000+01

Transmit Electrometer Output

W<CR>

V1 Ave: 1.2765e+00

Transmit Fil Status

F<CR>

Filament #1 on High Voltage on

Transmit Averaged Pressure

P<CR>

Pa: 1.23456e+0 Torr<CR>

Transmit RS-485 Address

A<CR>

Multidrop Address: 01<CR>

Transmit High Set point

H<CR>

Hi: 1.00000e+1 Torr<CR>

Transmit Low Set point

L<CR>

Lo: 1.00000e-2 Torr<CR>

Transmit Device Status

S<CR>

00044<CR>

Transmit Turnaround Delay

T<CR>

Comm Delay: 6<CR>

Transmit Selected Units

U<CR>

Torr<CR>

Transmit Software Version #

V<CR>

IGE-3000 Vxxxx

Transmit Filament F<CR>

f1<CR>

 

Transmit Emission Setting

I<CR>

Emission: 0.01 mA<CR> (note:

 

 

Filament must be on)

Transmit Degas Time

D<CR>

Remaining Degas Time: 54

 

 

minutes<CR>

8.4.Parameter Modification Commands

Command Description

Format

Sample Response

Modify High Setpoint

H={m.dd}E{+e}<CR> 1.00000e-9 to 9.99999e+9

Modify Low Setpoint

L={m.dd}E{-e}<CR>

1.00000e-9 to 9.99999e+9

Modify RS-485 Address

*{aa}A={aa}<CR>

1 To DF (Hexadecimal)

Modify Turnaround Delay

*{aa}T={dd}<CR>

0 To 255 (Decimal)

Modify Emission

E={m.d}E{-e}<CR>

* must be exact

Modify Degas Time

R={dd}<CR>

0 To 5 (Decimal)

Turn Gauge/Filament on F1 or F2

F1<CR>

NA

Turn Gauge/Filament off

F0<CR>

NA

Escape

<Esc>

Reset Command Buffer (ignore prior

 

 

Input)

IGE 300

Page 19 of 21

Page 19
Image 19
Teledyne IGE 3000 Interrogation Commands, Parameter Modification Commands, Command Description Format Sample Response