Portable Oxygen Monitor

Service Manual

 

 

 

Symptom

Why

What To Do

 

 

ventilate the patient, water

 

 

vapor actually dilutes the

 

 

gas. See Appendix:

 

 

Humidity, Temperature.

 

 

D) If a blender is used, check

 

 

its calibration. See

 

 

Appendix: Discrepancy in

 

 

Readings.

The unit has stopped

The MX300-I

A) Disconnect the batteries and

working and the LCD is

instruments are

inspect the contacts for

displaying

equipped with an

corrosion. Reconnect the

alphanumeric figures.

electronic "watch dog,"

batteries. If the unit functions

 

which monitors the

properly, calibrate the unit

 

circuitry within the unit

and reset the alarm values.

 

for potential faults and

B) Try a new set of batteries.

 

renders the unit

 

 

 

inoperable until the

C) Increase the distance

 

condition is corrected.

between the unit and any

 

(See Watch dog

source of radio frequency

 

section below)

interference. The sensor

 

Several conditions

cable is a prime source of

 

pickup as it can act like an

 

can activate the

 

antenna. Relocate the

 

"watch dog."

 

sensor cable and if possible

 

Dropping the unit,

 

change its coiled length to

 

poor battery

 

“de-tune” its antenna effect.

 

connections, and

 

Placing the cable in a

 

radio frequency

 

different position may also

 

interference are the

 

help.

 

most common

 

 

 

causes. See the

 

 

watch dog section for

 

 

additional information

 

Alarm sounds/flashes

A) Readings are

A) Adjust high and low alarm

continuously.

outside alarm

setting to be above and

 

limits.

below O2 value being

 

 

displayed.

No display.

A) Batteries expired.

A) Check/replace batteries.

 

B) Bad battery

B) Check battery connections.

 

connection.

C) Calibrate.

 

 

 

 

 

Keys inoperable/cannot

The LOCK/UNLOCK

A) Press LOCK/UNLOCK key

turn unit off

key is activated which

once. LCD will flash

 

is preventing key

indicating keys are active.

 

operation

 