Portable Oxygen Monitor | Operation | |
|
|
|
Operation
Note: Upon receipt, INSPECT THE ENTIRE UNIT FOR DAMAGE. Check the unit and all included accessories for broken or loose parts. If damaged, DO NOT USE. Notify the shipper, and consult Teledyne Analytical Instruments.
Note: This equipment is internally powered using 3 AA batteries.
CAUTION: THE
2.1 Setup
The
To set up and use your
1.Install the sensor.
2.Install the batteries.
3.Calibrate the unit.
4.Set the alarms.
The control keys are designed for easy operation. A LOCK/UNLOCK key has been supplied to prevent accidental changes to critical settings. This eliminates unwanted changes in calibration or alarm settings from accidental touching or bumping of the keys. To further reduce the possibly of incorrect adjustments at least two keys must be pressed in order to modify a critical calibration or alarm set point value.
Note: The ALARM SILENCE () and BATT TEST key continue to operate normally when the lock feature is activated.