Portable Oxygen Monitor

Service Manual

 

 

 

50

5

The ADC circuit failed during POST.

6 (5)

6

A stuck key has been detected. The second digit

 

(There is no

example (5) on the display shows which key is

 

stuck:

 

indication of

 

 

 

 

which key is

0 - Low Alarm

 

stuck)

1

- Batt

 

 

 

 

2

- Cal

 

 

3

- Up

 

 

4

- Down

 

 

5

- Silence

 

 

6

- High Alarm

 

 

7

- Alarm Test

 

 

8

- Key Lock

 

 

 

7(2)

7

A failure has occurred during the automatic

 

(There is no

calibration of the digital to analog converter (DAC)

 

circuit. The second digit example (2) shows the

 

indication of

 

type of digital to analog failure.

 

the type of

 

0 - Measurement

 

DAC failure)

 

 

1

- High Test

 

 

2

- Low Test

 

 

3

- Offset Cal

 

 

4

- Gain Cal

3.10 Other Problems with the Instrument

Most other problems arise from either mechanical damage from the instrument falling from a bench or table, or electronic component failure. In these units, repair or troubleshooting the PCB or individual component on the board is not feasible. It requires specialized test equipment and probes not generally available to the public. Under most circumstances a replacement of the entire PC Board is recommended. The instrument must be returned to the factory for PCB installation.

Occasionally, depending on the environment of use, keys can become stuck or function erratically due to contamination. Use a mild