Teledyne MX300-I manual Troubleshooting

Models: MX300-I

1 48
Download 48 pages 16.73 Kb
Page 37
Image 37

Portable Oxygen Monitor

Service Manual

 

 

 

investigation should be made until the discrepancy in readings is resolved. The troubleshooting section of this manual may provide additional assistance in locating the problem.

Note: The MSDS on this material is located in the Appendix. It is also available upon request through the Teledyne Environmental, Health and Safety Coordinator. Contact at

(626)934-1592

3.8Troubleshooting

The MX300-I oxygen monitor provides a variety of built-in safety features that prevents its use when a fault is detected. When a unit displays the message √ SENSOR and sounds the audible and visual alarm continuously, it is an indication of a faulty connection between the sensor and the unit or an expired or faulty sensor. To determine where the difficulty lies, refer to the following guidelines in Table 3-1.

Table 3-1 Troubleshooting

Symptom

Why

What To Do

New sensor responds

If the sensor is new

A) Wait 1–2 hours and

slow or drifts.

and was just

recalibrate.

 

removed from its

 

 

sealed bag it may

 

 

need to run for

 

 

several hours.

 

Sensor will not read

Calibration in 100%

A) Recalibrate using dry gas

below 22 % after

was invalid or the

making sure the reading

calibration in 100% O2.

room air is

stabilizes before making

 

contaminated with

any adjustments.

 

excess oxygen.

B) Make sure that at least 6"

 

 

 

 

(30 cm.) of tubing is

 

 

attached to the exhaust

 

 

side of the tee adapter to

 

 

prevent back filling. O2 flow

 

 

rate should not exceed 5

 

 

l/min.

 

 

C) Oxygen concentration at the

 

 

sensor is significantly higher

 

 

than 21%. Take the

 

 

instrument to a well-

 

 

ventilated area and check

Teledyne Analytical Instruments

37

Page 37
Image 37
Teledyne MX300-I manual Troubleshooting