The specific descriptions for each type of alarm are as follows:

Power Fail Alarm: This alarm changes state when the AC or DC system power is lost.

Alarm #1: Programmable as high or low alarm (actuates when concentration is above threshold). Can be set anywhere between 0 and 9999ppm, but must be set ABOVE the threshold set for Alarm 2.

Alarm #2: Programmable as high or low alarm (actuates when concentration is below threshold). Can be set anywhere from 0 to 9999ppm.

CAUTION: THERE COULD BE HAZARDOUS VOLTAGE AT THE ALARMS TERMINALS, EVEN WHEN POWER IS REMOVED FROM THE INSTRUMENT.

3.3.4 Relay Ratings

The 3300/OT-3 has 2 relay ratings.

Higher rating —established by the manufacturer

DC Rating: 0-30 VDC maximum @ 2.0 Amps (resistive)

AC Rating: 125 VAC maximum @ 0.5 Amps

minimum reliable switching rating: 10 μA @ 10mV

Lower rating—maximum safe rating for Cl 1 Div 2 Group C or D hazardous location

DC Rating: 0-30 VDC maximum @ 0.9 Amps maximum

AC Rating: 125 VAC maximum @ 0.2 Amps maximum

minimum reliable switching rating: 10 μA @ 10mV

To safely use the alarm relays in a hazardous location, the characteristics of both the switched voltage and the components or load being operated must be known. The alarm relays must not be attached to