Operation

OT-3

 

 

will cause the display to flash NO PASS and the keys will have no effect.

4.To unlock the keypad, repeat steps 1 and 2. Enter the number 10 when the word PASS CODE blinks on the display.

4.7Selecting a Fixed Range or Autoranging

The Model OT-3 can operate in fixed high, fixed low, or autoranging mode. These modes are user-selectable. To change modes:

1.Press and then release the SET HI RANGE and the SET LO RANGE buttons simultaneously.

2.Within 5 seconds, press either the D or Ñ button until Auto, Lo, or Hi displays on the LED, as desired.

After about three seconds, the analyzer resumes monitoring in the selected range mode.

Note: If the concentration exceeds 9999 ppm oxygen, the analyzer will automatically switch to the Calibration Range,

EVEN THOUGH INSTRUMENT IS IN THE FIXED RANGE MODE.

4.8Calibration

Preliminary—If not already done: Power up the Analyzer and allow the LED reading to stabilize. Set the Alarm setpoints and the full-scale ranges to the desired values.

Procedure:

Note: This procedure requires a suitable span gas with pressure regulation.

The O2 concentration of the span gas should be 80-90% of the principle range of use— typically 8-9 ppm or 80-90 ppm.

1.Attach the pressure regulated span gas to the to the span port on the bottom of the analyzer as described in Section 3.4.

2.Move the gas selector valve to the SPAN position.

3.Adjust the pressure on the span gas to produce a 0.4-0.5 SCFH flow and allow the analyzer to stabilize before setting the span.

4.The analyzer reading should be stable prior to setting the span.

Teledyne Analytical Instruments

28

Page 36
Image 36
Teledyne OT-3 operating instructions Selecting a Fixed Range or Autoranging, Calibration