Trace Oxygen Analyzer | Maintenance | |
|
|
|
CAUTION: READ THE SECTION CELL WARRANTY CONDITIONS, BELOW, BEFORE REPLACING THE CELL.
After replacing the
5.2.2 Ordering and Handling of Spare Sensors
To have a replacement cell available when it is needed, TAI recommends that one spare cell be purchased when the current cell 's warranty period is approximately two thirds over.
CAUTION: DO NOT STOCKPILE CELLS. THE WARRANTY PERIOD STARTS ON THE DAY OF SHIPMENT. FOR BEST RESULTS, DO NOT ORDER A NEW SPARE CELL TOO SOON.
The spare cell should be carefully stored in an area that is not subject to large variations in ambient temperature (75 °F nominal), and in such a way as to eliminate the possibility of incurring damage.
CAUTION: DO NOT DISTURB THE INTEGRITY OF THE CELL PACKAGE UNTIL THE CELL IS TO ACTUALLY BE USED. IF THE CELL PACKAGE IS PUNCTURED AND AIR IS PERMITTED TO ENTER,
WARNING: THE SENSOR USED IN THE MODEL
5.2.3 Removing the Micro-fuel Cell
To remove a spent or damaged
1. Disconnect the Power Source from the instrument.
Teledyne Analytical Instruments | 33 |