Trace Oxygen Analyzer

Operation

 

 

 

When the alarms are set for delay and an alarm condition exists, the display will alternate between the O2 reading and the time in minutes and seconds before the alarm will trigger. The alarm indicator on the front panel will blink on and off. The delay time will continue to count down until it reaches zero then the alarm relay will close. At this point the display will stop blinking and the O2 concentration will display continuously. If a second alarm is reached after the delay time has elapsed, it will trigger immediately without any delay. If the O2 level drops to below the alarm condition during the delay, the delay timer will reset to the full delay.

If the alarm delay is set to 0 and an alarm condition exists, the alarm indicator will blink and the alarm relay will close immediately. When the O2 level drops out of the alarm condition, the alarm relay will drop out accordingly.

The alarm delay can be toggled on or off and the timing set through the DELAY function. To set the alarm delay:

1.Press the UP and DOWN arrow keys simultaneously. The display will indicate either PASS CODE or DELAY. If the display indicates PASS CODE, then to set the delay, press both arrow keys simultaneously again to toggle to DELAY.

2.While the display is blinking the word PASS DELAY, press the UP arrow key to increase the delay. To decrease the delay, press the DOWN arrow key. After three seconds the delay feature will time-out and return to the analyze mode.

4.6Keypad Lockout

The Model OT-3 has a keypad lockout feature which is designed to prevent unauthorized changes to the analyzer. This feature is toggled on or off via the PASS CODE function. To enable the keypad lockout:

1.Press the UP and DOWN arrow keys simultaneously. The display will indicate either PASS CODE or DELAY. If the display indicates DELAY, then to set the lockout feature, press both arrow keys simultaneously again to toggle to PASS CODE.

2.While the display is blinking the word PASS CODE, press the up or down arrow key to alter the code number. Any number except 10 will lock the keypad.

3.After three seconds, the control unit will time-out and return to analyze mode. With the keypad locked, pressing the keys

Teledyne Analytical Instruments

27

Page 35
Image 35
Teledyne OT-3 operating instructions Keypad Lockout