A.1 Specifications

Ranges:

Two user selectable ranges can be set

 

between 0-10 ppm and 0-9999 ppm

 

oxygen, and a 0-25 % (nominal) Air

 

Calibration Range.

Signal Output:

Voltage:

0–10 VDC, negative ground

 

Current:

4-20 mA, negative ground

Range ID:

0-10 VDC.

 

Display:

Light emitting diode (LED) display.

Alarms:

Two user-settable non-latching alarms,

 

with user adjustable programmable 0-20

 

min.delay; one power failure relay. (All

 

are failsafe.)

System Operating Temp:

0-50 °C

 

Accuracy:

±2 % of full scale at constant temperature

 

±5 % of full scale through operating

 

temperature range (At 100 ppm and higher

 

user defined ranges) once temperature

 

equilibrium is reached.

 

±1 ppm for 10 ppm range under above

 

conditions.

 

Response Time:

90 % in less than 65 seconds at 25°C (68°F).

System Power Requirement:

 

 

AC (100-240 VAC, 50/60Hz @ 2.8W), or

 

DC (10-36 VDC @2.8W )

Weight:

30.9Lbs (14 Kg)

System Enclosure:

NEMA 4 enclosure.

 

40 cm x 30 cm x 20.6 cm

 

(15.8” x 11.8” x 8.1”)