Trace Oxygen Analyzer | Operation | |
|
|
|
4.3.1 HI Range
Setting the HI Range
1.Press the SET HI RANGE Function button once.
2.Immediately (within 5 seconds) press either the D or Ñ button to raise or lower the displayed value, as required, until the display reads the desired
4.3.2LO Range
Setting the LO Range
1.Press the SET LO RANGE Function button once.
2.Immediately (within 5 seconds) press either the UP or DOWN arrow button to raise or lower the displayed value, as required, until the display reads the desired
4.4Setting the Alarm Setpoints
The alarm setpoints can be adjusted over the full range of the analyzer
4.4.1 Set Alarm 1
Alarm 1 can be set either as a high or low alarm. To configure this alarm to your preferences:
1.Press the SET ALARM 1 function button once.
2.The display will flash either HI or LO depending on what the alarm was configured in the last time. If the alarm configuration must be changed, use the Up or Down keys to toggle the alarm from HI to LO or
3.To change the setting at which the alarm will be actuated, press the SET ALARM 1 function button once more. The alarm setpoint will flash on the LED display. Press either the Up or Down keys to raise or lower the displayed value, as required,