7. System Maintenance

7. System Maintenance

Waste Gas Cleaning System:

See drawing on page 42.

Attention:

Empty the condensate reservoir completely after each measuring

 

operation. Water residues within the measuring instrument will destroy the

 

pumps and sensors!

 

Damage of the filter and / or improperly fitted filter will greatly decrease o

 

eliminate the filter function and will eventually destroy pumps and sensors.

 

Check the microfilter for contaminations and replace as necessary.

 

If the pump capacity is reduced, exchange the diaphragm filter.

 

Make sure that threaded parts are straight when placed on and tighten them

 

moderately. Ensure sufficient sealing by means of O-rings.

Plug-type elements

 

and flanges:

Remove any gas residues. Grease with Vaseline.

Storage:

Store in a cool and dry environment at a temperature of approx. 60 °F (20 °C).