Input Screen

This screen allows you to change the input channel range and fullscale voltage (see the Principles section (5.2) for more information on this) and also assign a units label (up to 5, non-scaling characters) for the channel data. The number of decimal places selected on the range field dictates the number of decimal places shown for the input channel data on the Main screen. To reduce the number of decimal places, change the number under the edit cursor to a decimal point and the system will auto format the field when ‘Zero’ is pressed.

Switch Functions:

Func

Go back to Main Menu screen

Select previous character when in edit mode

Select next character when in edit mode

Move edit field selection up

Move edit field selection down

Zero

Go to edit mode for currently selected field

e)Filtering Screen

This screen allows the user to modify the band and buffer size of the adaptive filter. The filter only applies to the display and serial communications output readings. The band is represented as a percentage of FS and can be set between 0.00% (OFF) and 1.00% and also to ON. The buffer size allows settings between 0 and 6 seconds of filtering. For more information on the adaptive filter, see the Principles section (5.6)

Switch Functions:

Func

Go back to Main Menu screen

Select previous character when in edit mode

Select next character when in edit mode

Move edit field selection up

Move edit field selection down

Zero

Go to edit mode for currently selected field