2.Installation and Interconnections

2.1.Installation

The THCD-100 can be panel mounted, or used on a bench top and is a standard 96 x 48mm panel meter enclosure. Note that the instrument must always be situated in such a way as to enable adequate air circulation about the unit. If a transducer was supplied with the THCD-100 a suitable cable may have been included.

See Section 2.6 for a cut out detail for panel mounting.

2.2.Transducer socket Pinout – 15 Pin 'D' Type

Pin Number

Designation

5

Signal Common

6

Signal In (VDC)

7

Case Gnd

9

-15 VDC

11

+15 VDC

12

Valve Return

13

+24 VDC

14

Set Point Out (VDC)

All other pins

Not Connected

If the transducer is being powered from another source (e.g. a mains powered supply) it will only be necessary to connect to the two signal pins. Note: The THCD-100 has a fully floating differential input. For single ended use (i.e. transducers with differential output) join 0V (pin5) & Signal Return (pin12) at the transducer. Ensure the common mode input range is not exceeded.