Revision A

 

 

 

PROBLEM

PROBABLE CAUSE

SUGGESTION

 

 

 

Recurrence of the first

The GoodKnight 420G is

Ask your home care provider to check the

symptoms of sleep

not set to the correct

operation of the GK420G.

apnea syndrome.

pressure or is not operating

 

 

correctly.

 

 

 

Your physical condition or

Contact the ordering physician.

 

your pressure

needs have

 

 

changed.

 

 

The GoodKnight 420G

Verify that the power

Check the connections between the power

does not come on (no

supply module is properly

supply module, the GK420G and the electric

display).

connected.

 

outlet.

 

No AC power.

 

Use another appliance, such as a lamp or radio

 

 

 

to check that the AC power supply is present at

 

The fuse of the device has

the outlet.

 

Contact your home care provider to get

 

blown.

 

information on the way of returning the device

 

 

 

to the technical department.

The GoodKnight 420G

Excessive electro-magnetic

Move the device away from sources of

is not running properly

interference.

 

interference such as halogen lamps and cell

and seems to be subject

 

 

phones, etc.

to interference.

 

 

 

9:33 is displayed.

The appliance is connected

This message disappears if you press the

(value given as a guide only)

to the mains and remains

ON/OFF button. (See reference 4 in Figure 2,

 

on standby.

 

page 5).

4.0 cmH2O is

Set pressure level

This pressure level has been prescribed by your

displayed.

(equivalent to the pressure

doctor; it can only be altered by medical

(value given as a guide only)

in the mask for a hose of

prescription.

 

1.80 m).

 

If the ramp function is active, this level of

 

 

 

pressure will increase progressively before

 

 

 

reaching the prescribed value and the ramp

 

 

 

indicator is displayed.

In 01 is displayed.

The mask is disconnected.

Check the connections between the mask, the

 

 

 

hose and the device. This message disappears

 

 

 

as soon as the mask is well connected.

In 03 is displayed.

Reduction of

the power

Check the connections between the power

 

supply voltage.

 

supply module, the GK420G and the mains

 

 

 

socket. Unplug the power supply module, then

 

 

 

connect it again to the mains. If the problem

 

 

 

remains, contact the home care provider.

 

 

 

Check the external battery and replace it, if

 

 

 

necessary. If the problem remains, contact the

 

 

 

home care provider.

Er XX appears on the

The device has detected an

Contact the home care provider.

display.

operating fault.

 

 

(XX = 2 digits).

 

 

 

20

GoodKnight 420G - Patient Manual