Class II power supply:

Revision C

Electrical characteristics of the power supply module

Input voltage: 100 – 240 VAC (-10%, +10%), 50 / 60 Hz (±1 Hz) Input current: 550 mA (Egston), 600 mA (Anam).

Output voltage: 13 V – 1.85 A (Egston) / 12 V – 2 A (Anam).

Power supply provided with the device: Egston N2UFMW 24 W 13 V or Anam TA2412 24 W 12 V.

CAUTION

Use only one of the above-mentioned specific power supplies provided with the device.

The above-mentioned power supply modules are not intended to be repaired. In case of failure, please contact your service provider to have the module replaced.

Electrical characteristics of the device

Input voltage:

13 V – 1.5 A.

Maximum power consumed : 20 W (without humidifier)

Current consumed at 20 cmH2O with a 4 mm leak: 0.750 A.

Direct current

13 V power connector

Symbols

0459

ON/OFF

Information Access

Ramp

Increases setting values

Reduces setting values

Output connector

Device protected against the vertical falls of water drops.

Type B device

Consult the instructions for use before using this device.

Device complies with the requirements of the directive 93/42/CEE relating to medical appliances.

For USA only:

Federal law restricts this device for sale by or on the order of a physician.

22

GoodKnight 420G - Patient Manual