Revision B

TECHNICAL CHARACTERISTICS

Device performance

Pressure range of the GoodKnight 420G:

Maximum achievable pressure at the patient connection port under single fault condition:

Ramp time:

Sound level measured in accordance with: PrEN ISO 17510-1 :2001

Maximum sound level:

Lifetime of the battery soldered onto the board:

4cmH2O to 20 cmH2O ± 1 cmH2O adjustable by increments of 0.5 cmH2O

24 cmH2O

0 to 30 minutes ± 1 minute

device adjustable by 5-minute increments 25.1 dBA

31dBA at 10 cmH2O > 5 years

Physical characteristics

Dimensions (W x H x L ):

5.6 x 2.9 x 7.7 in (without power supply)

Weight:

1.54 lb (without power supply module)

Transport and storage conditions

Temperature:

-20°C to +60°C (-4°F to +140°F)

Relative humidity:

up to 95 % without condensation

Relative pressure range:

500 hPa to 1060 hPa

 

(1 hecto-Pascal 1 cmH2O)

Conditions of use

Temperature:

+5°C to +40°C (+41°F to +104°F)

Relative humidity:

between 10 % and 95 % without condensation

Relative pressure range:

700 hPa to 1060 hPa

 

(1 hecto-Pascal 1 cmH2O)

Operating Altitude:

0 to approximately 7,800 ft (0-2400 m)

Disposal of the GoodKnight 420G at the end of its life

In order to protect the environment, various parts of the equipment and its accessories must be disposed of in conformity with the local regulations.

GoodKnight 420G - Patient Manual

21