Kodak 750, 560 manual Troubleshooting, Symptom CAUSE/REMEDY

Page 30

Troubleshooting

SYMPTOM

CAUSE/REMEDY

Power is not supplied.

Is the power cord connected? Replace the power cord.

 

 

Is there a connection to the main switch?

 

Replace the main switch: P/N 1A154A084-01 (560 or 750 Series).

Circuit breaker CB1 trips when

There may be a short circuit of the fan motor or transformer T-10.

 

main switch is turned to LAMP

Replace the defective part (depending on your model) :

position.

P/N 1A166A106-03

 

 

P/N 1A669A140-01 (M4)

 

P/N 1A6699A107-01 (M3)

 

P/N 1A165F033-06 (M2)

 

P/N 1A165F039-01 (M5, M6)

 

P/N 1A165F033-06 (M1)

 

P/N 1A164A101-01 (T-10)

Xenon lamp will not start (starter

Is the lamp positive wire too close to the metal of the body frame?

 

spark is detected when the main

Change the position of the lamp wire. It should be at least 15mm from

switch is turned to the LAMP

all metal.

position).

There may be a defective Xenon lamp. This is one indication that the

 

 

lamp life has expired.

 

Replace the lamp (P/N XB-560 or P/N XB-750, depending on your

 

model).

No starter spark is detected when

85V DC may be across the starter or the starter may be defective.

 

the main switch is turned to the

Replace the starter with P/N 1A685A088-01 (560 Series, no housing), P/N

LAMP position.

1A159A035-01 (560 Series, black housing) or P/N 1A817S006-01 (750

 

Series).

 

70V or less is across the starter, +12V is not present between pins 15 &

 

14 of IC Q1.

 

Check the connector: CN12 P/N 1A172A058-03 (560 Series) or CN52

 

P/N 1A172B070-03 (750 Series).

Dark Xenon lamp.

The lamp current may be under the current of 25 Amps for the 560 Series

 

 

or under 37.5 Amps for the 750 Series.

 

Adjust VR11 of pulse width PCB for the 560 Series or VR1 of pulse width

 

PCB for the 750 Series.

 

The lamp may be worn out.

 

Replace the lamp (P/N XB-560 or P/N XB-750, depending on your

 

model).

Dark picture or uneven illumination

The lamp or optics may be improperly positioned.

 

of image.

Adjust the lamp position and check alignment of reflectors and the

 

reflector holders.

 

The Xenon lamp may be defective.

 

Replace the lamp (P/N XB-560 or P/N XB-750, depending on your

 

model).

 

NOTE: DON’T BREAK THE PAINT ON THE REFLECTOR MOUNT

 

SCREWS TO CHECK IF THEY ARE LOOSE. A COMMON PROBLEM IS

 

THAT THE REFLECTOR HAS BECOME LOOSE FROM REPEATED

 

SHIPPING.

The biggest problem that will occur with these units is failure to perform preventative maintenance. After a unit has been shipped a number of times it is a good idea to check the electrical connections and to look for loose screws. If loose screws are found and tightened, you will save many problems later.

28

Image 30
Contents Navitar Xenon 560/750 High Intensity Slide Projector World Leader in Image QualityImportant Before Opening the BOX Containing Table of Contents Page Introduction Supplied Components Servicing & Repair560/750 System Diagrams Top ViewSide View Front View Rear ViewFeatures Features Installing a New Lamp Getting StartedImportant Before Opening the BOX Containing Installing a New Lamp Installing a New Lamp Arc Correcting Magnet 750W Operating the Unit & Remote Control Power Supply ProjectionCleaning the Mirrors & Optical System Remote ControlEmitter Adjusting the Lamp Position Lamp AdjustmentAdjusting the Lamp Position Slide Burning & Picture Flicker Explanation & RemedyMagnet Adjustment Solution Slide Moisture Problems Areas of High Humidity Simple SolutionsConnecting the Dissolve Interface to the Dissolve Module Units with Dissolve Module & Dissolve InterfaceAttaching a New Dissolve Module Dissolve Shutter Manual Knob Units with Dissolve Module & Dissolve Interface Safety Precautions Lamp Location of Components SeriesLocation of Components Troubleshooting Symptom CAUSE/REMEDYSchematic Diagrams 560AF/DM750AF/DM 560 & 750 RAK/RARS232 Specifications Optical System Power Supply Slide ProjectorSpecifications Lamp Specifications General Components Specifications Features Navitar Dissolve Interface MDC200INL Type Imput Connector Fuse AC In Connector Operation Safety Precautions Examples of Connections Warranty Year Parts Year Labor