Installation and Configuration

Remote/Local Operation

If the rear panel PON REM switch is set to 1, the power supply will power up in local mode. Power supply control is at the front panel. During an operating session, you can toggle between local mode and remote mode by sending commands as described in “Remote Mode Operation” on page 22 and “Local Mode Operation” on page 23.

Table 2.5 Remote Mode Power On Conditions

Condition

Default Settings

7.5-140 Model Example

 

 

 

Voltage

0 V

VSET 0

 

 

 

Current

0 A

ISET 0

 

 

 

Soft Voltage Limit

VMAX (see models)

VMAX 7.5

 

 

 

Soft Current Limit

IMAX (see models)

IMAX 140

 

 

 

OVP Trip Voltage

Model VMAX + 10%

OVSET 8.25

 

 

 

Delay

0.5 s

DLY 0.5S

 

 

 

Foldback Protection

OFF

FOLD OFF

 

 

 

Output

ON

OUT ON

 

 

 

Hold

OFF

HOLD OFF

 

 

 

Unmask

NONE

UNMASK NONE

 

 

 

Service Request Capability

OFF

SRQ OFF

 

 

 

AUXA

OFF

AUXA OFF

 

 

 

AUXB

OFF

AUXB OFF

 

 

 

Remote Mode While in remote mode, use the interface commands to control the output of the Operation power supply from a computer. See Section 3, “Operation” for a complete list of device-dependent commands available with this interface. For information about how to use GPIB commands and GPIB lines, refer to the documentation for the

particular GPIB controller interface installed on your computer.

While in remote mode, you can change to local mode operation by sending the

GPIB GTL (Go to Local) command, or by negating the GPIB REN (Remote

Enable) line.

22

Operating Manual for GPIB for XPD Series Power Supply

Image 24
Contents GPIB-XPD Page Internal Gpib Interface for XPD 500 Watt Series How do you get service? Phone 604 422 Toll Free North America 800 667 Fax 604 421What does this warranty not cover? Disclaimer Product Information About Your Power Supply Model Number Serial Number Purchased From Purchase DateSafety SupplyInstallation and Configuration Main SectionsManual Revisions About This Manual Contents Contents Features and Specifications DescriptionFeatures and Functions Features and SpecificationsFunctions ReadbackSpecifications Models 18-30 33-16 60-9 120-4.5OVP Features and Specifications Introduction Initial InspectionPower Supply Front Panel with Gpib Interface Installed Gpib Interface Subplate Located on Power Supply Rear Panel Gpib Interface PCB Basic Setup Procedure Step # Description Action ReferencePON SRQ IEEE-488 Primary Address Selection AddressPower On Service Request PON SRQ Rear Panel SRQ Switch Position PON SRQ StateSwitch Position Power on results Condition Default Settings Model Example IsetIEEE-488 Controller Connection Internal PCB Jumper SelectionsOVP Control Mode Selection User Signals High OFF LOWHigh LOW OFF User Signals J21 Connector J21 User Signal Connector Circuit Block Diagram J21 User Cable with Ferrite Block Operation Gpib OperationExample Command directed to entire bus, or Bit Decimal Description Reset By Position Weight Request Service SRQ Set whenPower On PON Set when unit LSBCommand Syntax Computer EntryFormat Description Form Floating Number Definition ExampleIset 2.0A Vset Command Summary AUXA? AUXB?CMODE? DLY?FAULT? CmodeIdata IHICommand Reference Command DescriptionAuxa 1 on Auxb 1 onChanging an output setting Response Fold mode where mode is Snsp Mask None = Unmask ALLStatus, and Fault Registers on Accumulated Status, Status, and Fault Registers on SNSP, ALL, NoneResponse Vmax voltage Accumulated Status, Status, and Fault Registers Condition Mnemonic Bit Position Bit WeightError Codes 11 Error CodesTroubleshooting LEDsCalibration Voltage Mode Calibration SetupVoltage Program Calibration Procedure Readback VoltageCalibration Current Mode Calibration Current Calibration SetupCurrent Program Calibration Procedure Current Readback Calibration Procedure Over Voltage Protection OVP Calibration Page Xantrex Technology Inc