Accumulated Status, Status, and Fault Registers

Accumulated Status, Status, and Fault Registers

The GPIB option card uses three separate registers which are always active. They are the accumulated status, status, and fault registers. You can use the status commands shown in Table 3.8, “Status Commands” to activate the registers. The bit register has twelve conditions, each assigned a bit weight. When querying a register, the controller returns a response which is the sum of the weights of all relevant conditions.

Example:

ASTS?

Query the Accumulated Status register.

 

ASTS <771>

Controller response.

 

 

771 = 512 + 256 + 2 + 1 = PON + REM + CC + CV

 

 

Table 3.10 Accumulated Status, Status, and Fault Registers

 

Condition

 

Mnemonic

Bit Position

Bit Weight

 

 

 

 

 

Constant voltage operation

 

CV

0

1

 

 

 

 

 

Constant current operation

 

CC

1

2

 

 

 

 

 

Not used

 

2

4

 

 

 

 

 

Overvoltage protection tripped

 

OV

3

8

 

 

 

 

Over temperature protection tripped

OT

4

16

 

 

 

 

Supply external shutdown active (J21-1)

SD

5

32

 

 

 

 

 

Foldback mode operation

 

FOLD

6

64

 

 

 

 

 

Remote programming error

 

ERR

7

128

 

 

 

 

Power ON (accumulated status, status

PON

8

256

registers only)

 

 

 

 

Remote mode (accumulated status, status

REM

9

512

registers only)

 

 

 

 

AC fail condition

 

ACF

10

1024

 

 

 

 

 

Output fail condition

 

OPF

11

2048

 

 

 

 

 

Sense protection tripped

 

SNSP

12

4096

Notes:

1.All mnemonics can be masked or unmasked.

2.The error (ERR) bit is reset in the accumulated status, status, and serial poll registers with an error query (ERR?).

3.The accumulated status register is cleared with an accumulated status query (ASTS?).

4.A fault is cleared with a fault query (FAULT?).

46

Operating Manual for GPIB for XPD Series Power Supply

Image 48
Contents GPIB-XPD Page Internal Gpib Interface for XPD 500 Watt Series How do you get service? Phone 604 422 Toll Free North America 800 667 Fax 604 421What does this warranty not cover? Disclaimer Product Information About Your Power Supply Model Number Serial Number Purchased From Purchase DateSafety SupplyInstallation and Configuration Main SectionsManual Revisions About This Manual Contents Contents Features and Specifications DescriptionFeatures and Functions Features and SpecificationsFunctions ReadbackSpecifications Models 18-30 33-16 60-9 120-4.5OVP Features and Specifications Introduction Initial InspectionPower Supply Front Panel with Gpib Interface Installed Gpib Interface Subplate Located on Power Supply Rear Panel Gpib Interface PCB Basic Setup Procedure Step # Description Action ReferencePON SRQ IEEE-488 Primary Address Selection AddressPower On Service Request PON SRQ Rear Panel SRQ Switch Position PON SRQ StateSwitch Position Power on results Condition Default Settings Model Example IsetIEEE-488 Controller Connection Internal PCB Jumper SelectionsOVP Control Mode Selection User Signals High OFF LOWHigh LOW OFF User Signals J21 Connector J21 User Signal Connector Circuit Block Diagram J21 User Cable with Ferrite Block Operation Gpib OperationExample Command directed to entire bus, or Bit Decimal Description Reset By Position Weight Request Service SRQ Set whenPower On PON Set when unit LSBCommand Syntax Computer EntryFormat Description Form Floating Number Definition ExampleIset 2.0A Vset Command Summary AUXA? AUXB?CMODE? DLY?FAULT? CmodeIdata IHICommand Reference Command DescriptionAuxa 1 on Auxb 1 onChanging an output setting Response Fold mode where mode is Snsp Mask None = Unmask ALLStatus, and Fault Registers on Accumulated Status, Status, and Fault Registers on SNSP, ALL, NoneResponse Vmax voltage Accumulated Status, Status, and Fault Registers Condition Mnemonic Bit Position Bit WeightError Codes 11 Error CodesTroubleshooting LEDsCalibration Voltage Mode Calibration SetupVoltage Program Calibration Procedure Readback VoltageCalibration Current Mode Calibration Current Calibration SetupCurrent Program Calibration Procedure Current Readback Calibration Procedure Over Voltage Protection OVP Calibration Page Xantrex Technology Inc