Xantrex Technology GPIB-XPD manual Current Readback Calibration Procedure

Page 55

Current

Readback

Calibration

Procedure

Calibration

Current Mode Calibration

1.Connect the current shunt and voltmeter to the power supply as shown in Figure 4.2.

2.Activate calibration mode by sending command CMODE ON or CMODE 1 to the power supply.

3.Send command IRLO; VRLO to the power supply. Wait for the supply to settle. Measure and record the output voltage shown on the external voltmeter. Send IRLO again.

4.Send command IRHI; VRHI to the supply. Wait for the supply to settle. Measure and record the output voltage shown on the external voltmeter. Send IRHI again.

5.Calculate IRLO and IRHI from the voltages taken from the external voltmeter and the shunt resistance. I=V/R.

6.Send the command IRDAT <ilo>,<ihi> to the power supply. <ilo> and <ihi> are the current values obtained from sending the IRLO and IRHI commands to the power supply. When the power supply is calibrated, the low to high current readback calibration values are stored as constants.

7.Program the supply at various levels using the ISET command to confirm that the calibration was successful and that linearity is observed. Refer to the current readback accuracy specification in Section 1.

8.Turn off calibration mode by sending the command CMODE OFF or CMODE 0 to the power supply.

Release 1.2

53

Image 55
Contents GPIB-XPD Page Internal Gpib Interface for XPD 500 Watt Series Phone 604 422 Toll Free North America 800 667 Fax 604 421 How do you get service?What does this warranty not cover? Disclaimer Product Model Number Serial Number Purchased From Purchase Date Information About Your Power SupplySupply SafetyMain Sections Installation and ConfigurationManual Revisions About This Manual Contents Contents Description Features and SpecificationsReadback Features and FunctionsFeatures and Specifications FunctionsModels 18-30 33-16 60-9 120-4.5 SpecificationsOVP Features and Specifications Initial Inspection IntroductionPower Supply Front Panel with Gpib Interface Installed Gpib Interface Subplate Located on Power Supply Rear Panel Gpib Interface PCB Step # Description Action Reference Basic Setup ProcedurePON SRQ Address IEEE-488 Primary Address SelectionRear Panel SRQ Switch Position PON SRQ State Power On Service Request PON SRQSwitch Position Power on results Iset Condition Default Settings Model ExampleInternal PCB Jumper Selections IEEE-488 Controller ConnectionOVP Control Mode Selection High OFF LOW User SignalsHigh LOW OFF User Signals J21 Connector J21 User Signal Connector Circuit Block Diagram J21 User Cable with Ferrite Block Gpib Operation OperationExample Command directed to entire bus, or LSB Bit Decimal Description Reset By Position WeightRequest Service SRQ Set when Power On PON Set when unitComputer Entry Command SyntaxFormat Floating Number Definition Example Description FormIset 2.0A Vset Command Summary DLY? AUXA?AUXB? CMODE?IHI FAULT?Cmode IdataAuxb 1 on Command ReferenceCommand Description Auxa 1 onChanging an output setting Response Fold mode where mode is Mask None = Unmask ALL SnspStatus, and Fault Registers on SNSP, ALL, None Accumulated Status, Status, and Fault Registers onResponse Vmax voltage Condition Mnemonic Bit Position Bit Weight Accumulated Status, Status, and Fault Registers11 Error Codes Error CodesLEDs TroubleshootingCalibration Voltage Voltage Mode CalibrationSetup Voltage Program Calibration Procedure ReadbackCalibration Current Calibration Setup Current Mode CalibrationCurrent Program Calibration Procedure Current Readback Calibration Procedure Over Voltage Protection OVP Calibration Page Xantrex Technology Inc