Xantrex Technology GPIB-XPD manual Calibration

Page 53

Calibration

Voltage Mode Calibration

3.Send command VRLO; IRLO to the power supply. Wait for the supply to settle. Measure and record the output shown on the external voltmeter. Send VRLO again.

4.Send VRHI; IRHI to the supply. Wait for the supply to settle. Measure and record the output voltage shown on the external voltmeter. Send VRHI again.

5.Send the command VRDAT <vlo>,<vhi> where <vlo> and <vhi> are the values read from the voltmeter after the VRLO and VRHI commands were sent. The processor calculates the offset value required to calibrate the power supply.

When the power supply is calibrated, the low to high voltage readback calibration values (offsets) are stored as constants.

6.Use commands VSET and VOUT? commands to confirm that the calibration was successful and that linearity is observed. Refer to the voltage readback accuracy specification in Section 1.

7.Turn off calibration mode by sending the command CMODE OFF or CMODE 0 to the power supply.

Release 1.2

51

Image 53
Contents GPIB-XPD Page Internal Gpib Interface for XPD 500 Watt Series Phone 604 422 Toll Free North America 800 667 Fax 604 421 How do you get service?What does this warranty not cover? Disclaimer Product Model Number Serial Number Purchased From Purchase Date Information About Your Power SupplySupply SafetyManual Revisions Installation and ConfigurationMain Sections About This Manual Contents Contents Description Features and SpecificationsFeatures and Specifications Features and FunctionsFunctions ReadbackOVP SpecificationsModels 18-30 33-16 60-9 120-4.5 Features and Specifications Initial Inspection IntroductionPower Supply Front Panel with Gpib Interface Installed Gpib Interface Subplate Located on Power Supply Rear Panel Gpib Interface PCB PON SRQ Basic Setup ProcedureStep # Description Action Reference Address IEEE-488 Primary Address SelectionSwitch Position Power on results Power On Service Request PON SRQRear Panel SRQ Switch Position PON SRQ State Iset Condition Default Settings Model ExampleInternal PCB Jumper Selections IEEE-488 Controller ConnectionOVP Control Mode Selection High LOW OFF User SignalsHigh OFF LOW User Signals J21 Connector J21 User Signal Connector Circuit Block Diagram J21 User Cable with Ferrite Block Gpib Operation OperationExample Command directed to entire bus, or Request Service SRQ Set when Bit Decimal Description Reset By Position WeightPower On PON Set when unit LSBFormat Command SyntaxComputer Entry Floating Number Definition Example Description FormIset 2.0A Vset Command Summary AUXB? AUXA?CMODE? DLY?Cmode FAULT?Idata IHICommand Description Command ReferenceAuxa 1 on Auxb 1 onChanging an output setting Response Fold mode where mode is Mask None = Unmask ALL SnspStatus, and Fault Registers on SNSP, ALL, None Accumulated Status, Status, and Fault Registers onResponse Vmax voltage Condition Mnemonic Bit Position Bit Weight Accumulated Status, Status, and Fault Registers11 Error Codes Error Codes LEDs TroubleshootingCalibration Setup Voltage Mode CalibrationVoltage Program Calibration Procedure Readback VoltageCalibration Current Program Calibration Procedure Current Mode CalibrationCurrent Calibration Setup Current Readback Calibration Procedure Over Voltage Protection OVP Calibration Page Xantrex Technology Inc