Table 1.1 Specifications for XPD 500 W Series Supply with GPIB Interface Installed

Models

7.5-67

18-30

33-16

60-9

120-4.5

Program Resolution

 

 

 

 

 

Voltage

1.2mV

4.6mV

5.1mV

9.3mV

18.6mV

Current

5.2mA

3.6mA

2.9mA

1.3mA

0.7mA

OVP

1.2mV

4.6mV

5.1mV

9.3mV

18.6mV

Program Accuracy1

 

 

 

 

 

Voltage

10mV

75mV

75mV

150mV

180mV

 

±0.12%

±0.12%

±0.12%

±0.3%

±0.25%

Current

250mA

140mA

115mA

80mA

80mA

 

±0.1%

±0.1%

±0.15%

±0.15%

±0.1%

OVP

80mV

200mV

330mV

600mV

1.2V

Readback Resolution

 

 

 

 

 

Voltage

1.2mV

4.6mV

5.1mV

9.3mV

18.6mV

Current

5.2mA

3.6mA

2.4mA

1.3mA

0.7mA

Readback Accuracy1

 

 

 

 

 

Voltage

30mV

75mV

75mV

150mV

180mV

 

±0.12%

±0.12%

±0.2%

±0.3%

±0.25%

Current

250mA

140mA

115mA

80mA

80mA

 

±0.1%

±0.1%

±0.15%

±0.15%

±0.1%

1.Apply accuracy specifications according to the following voltage program accuracy example: Set a model XPD 18-30 power supply to 10 volts.

The expected result will be within the range of 10 volts ± 75mV ± 0.12% of the set voltage of 10 volts.

Release 1.2

13

Image 15
Contents GPIB-XPD Page Internal Gpib Interface for XPD 500 Watt Series Phone 604 422 Toll Free North America 800 667 Fax 604 421 How do you get service?What does this warranty not cover? Disclaimer Product Model Number Serial Number Purchased From Purchase Date Information About Your Power SupplySupply SafetyInstallation and Configuration Main SectionsManual Revisions About This Manual Contents Contents Description Features and SpecificationsReadback Features and FunctionsFeatures and Specifications FunctionsSpecifications Models 18-30 33-16 60-9 120-4.5OVP Features and Specifications Initial Inspection IntroductionPower Supply Front Panel with Gpib Interface Installed Gpib Interface Subplate Located on Power Supply Rear Panel Gpib Interface PCB Basic Setup Procedure Step # Description Action ReferencePON SRQ Address IEEE-488 Primary Address SelectionPower On Service Request PON SRQ Rear Panel SRQ Switch Position PON SRQ StateSwitch Position Power on results Iset Condition Default Settings Model ExampleInternal PCB Jumper Selections IEEE-488 Controller ConnectionOVP Control Mode Selection User Signals High OFF LOWHigh LOW OFF User Signals J21 Connector J21 User Signal Connector Circuit Block Diagram J21 User Cable with Ferrite Block Gpib Operation OperationExample Command directed to entire bus, or LSB Bit Decimal Description Reset By Position WeightRequest Service SRQ Set when Power On PON Set when unitCommand Syntax Computer EntryFormat Floating Number Definition Example Description FormIset 2.0A Vset Command Summary DLY? AUXA?AUXB? CMODE?IHI FAULT?Cmode IdataAuxb 1 on Command ReferenceCommand Description Auxa 1 onChanging an output setting Response Fold mode where mode is Mask None = Unmask ALL SnspStatus, and Fault Registers on SNSP, ALL, None Accumulated Status, Status, and Fault Registers onResponse Vmax voltage Condition Mnemonic Bit Position Bit Weight Accumulated Status, Status, and Fault Registers11 Error Codes Error CodesLEDs TroubleshootingCalibration Voltage Voltage Mode CalibrationSetup Voltage Program Calibration Procedure ReadbackCalibration Current Mode Calibration Current Calibration SetupCurrent Program Calibration Procedure Current Readback Calibration Procedure Over Voltage Protection OVP Calibration Page Xantrex Technology Inc