Calibration info | Jumps to the calculation information |
| submenu, which displays information |
| about the values taken into account to |
| recalculate the DICOM DF. |
|
|
READING ROOM DEFINITION SUBMENU
Measured ambient | Indicates the ambient light actually |
light | measured by the ambient light sensor on |
| the front of the display. This is a |
| value. |
|
|
Averaged ambient | Shows the average of the measured |
light | ambient light since the display was |
| switched on. This is a |
|
|
Reading room | Indicates the reading room type you are |
| editing. You can select another room to |
| edit by using the Up or Down touch |
| keys or rotating the control wheel. |
|
|
Max. ambient light | This indicates the maximum ambient light |
| that corresponds to the selected reading |
| room. |
| If the measured ambient light is higher |
| than the value entered here, you should |
| take measures to darken the room or |
| select another reading room type. |
|
|
Preset corr. value | This value is taken into account in the |
| calculation of the DICOM DF if Continuous |
| DICOM ALC is switched off and DICOM |
| Offset is set to “Preset”. |
| To each reading room type corresponds |
| another preset value. |
|
|