TO SELECT A READING ROOM SETTING:
1Select the DICOM display function. See To select a display function (DF):, on page 43.
2Switch Continuous DICOM ALC off. See To switch continuous DICOM ALC on/off:, on page 44.
3In the ALC & DICOM Options menu, select Reading Room.
4Select the reading room that corresponds to the room type the display is installed in.
5Return to the menu.
You can select the following reading room types:
DRR CR/DR/MAMMO | Corresponds to light conditions in |
| diagnostic reading rooms for computed |
| radiology, digital radiology or |
| mammography. This setting has the lowest |
| maximum ambient light. |
|
|
DRR CT/MR/NM | Corresponds to light conditions in |
| diagnostic reading rooms for computed |
| tomography, magnetic resonance or |
| nuclear medicine scans. |
|
|
Staff Office | Corresponds to light conditions in office |
| rooms. |
|
|
Clinical Viewing | Corresponds to light conditions in |
| diagnostic reading rooms for clinical |
| viewing. |
|
|
Emergency Room | Corresponds to light conditions in |
| emergency rooms. |
|
|
Operating Room | Corresponds to light conditions in operating |
| rooms. This setting has the highest |
| maximum ambient light. |
|
|