Planar C3i/EX manual Safety precautions, Safety tips

Page 6

Safety precautions

External equipment intended for connection to signal input, signal output, or other connectors, must comply with the relevant IEC standard (EN/IEC 60601-1 series for medical electrical equipment). In addition, all such combinations (systems) must comply with the standard IEC 60601-1-1, Safety requirements for medical electrical systems.

Equipment not complying to IEC 60601 must be kept outside the patient environment, as defined in the standard as at least 1.5 meters from the patient or the patient support.

Any person who connects external equipment to signal input, signal output, or other connectors has formed a system and is therefore responsible for the system to comply with the requirements of IEC 60601-1-1. If in doubt, speak

with a qualified technician.

Safety tips

Never open the display case, even when the power is off. Dangerous voltage inside may cause electric shock or death.

To avoid damage to the display, use the grounded power supply and video cable supplied by Planar, or use certified replacements.

Be sure the display is electrically grounded. You must connect the third grounding pin on the US power cord to a grounded outlet. The European power cord does not have a third grounding pin, but it must be plugged into a grounded outlet.

If you cannot insert the plug into the outlet you plan to use, have a licensed electrician replace the outlet with a properly grounded outlet. If the power cord connects directly into the computer, make sure the computer is grounded.

Keep the display dry if it is part of a surgical system. The display lacks protection against liquids or spills.

In locations where 240V outlets are used, connect the Dome display only to a center-tapped, 240V, single-phase supply (only for Canada and the United States).

GROUNDING RELIABILITY CAN ONLY BE ACHIEVED WHEN EQUIPMENT IS CONNECTED TO AN EQUIVALENT RECEPTABLE MARKED “HOSPITAL ONLY” OR “HOSPITAL GRADE.”

vi Dome Display

Image 6 Contents
User’s Guide 020-0377-01A Contents Product Information Symbol explanations Intended useSafety precautions Safety tipsUnpacking and handling tips Preventing fire and injuryCleaning the display Shipping/storing the display Disposal informationDVI cable About the Display System RequirementsDisplay Components Display Positions Tilt range Height rangeDesk Stand Features Stand cable coverInstalling the Display Moving from Dome C3i to Dome EX Display SystemTo enable Eprom Set the DIP Switch on the Dome DX2 BoardTo disable VGA To set landscape mode as default display modeInstall the Display Controller To install the display controllerConnect the Video and Power Cables To connect the cablesInstall Dome DX2 Driver and Dome CXtra Software Installation preparationRequired Sun software packages To install the driver and Dome CXtra Installing the Display Configure Display Settings Overriding the default display mode To edit the OWconfig fileDome Display To modify the OWconfig file Using visual classes To check the default visual To emulate a PseudoColor visual classSunallplanes Shape MIT-SHM Look Default visual id For visual Dome Display Configure the Window Environment Configuring the Common Desktop EnvironmentRunning OpenWindow from the console To use one or more Dome devices To use a Dome device with a Sun consoleDpms Screen Saver Screen saver tipsTo set the Dpms Screen Saver To use the setenv command Console ModificationDome CXtra Software Appendix Troubleshooting Problem Possible Cause SolutionDome C3i Specification Category Characteristic Item SpecificationPower Supply ReliabilityEnvironment Dome E2 Specification Power Supply Dome E3 Specification EMI shielding No emission of low-level radiation Temperature Dome E3c Specification Display Mtbf 30,000 hours Backlight Dome E5 Specification Display Mtbf 50,000 hours Backlight Video Modes 2560 xConnector Ports Details of DIN connectorsLED Status Lights Power-up sequence information from LED Index 21-22 Regulatory Compliance Canada, European Union, United StatesPage Fax + 358 9 420 EU Declaration of Conformity for Medical Application FCC Compliance Statement Japan AustralianChina KoreaStandard Warranty Standard Warranty Return ProcedureSummary Summary Limitations and Exclusions of Dome Displays Limitation of Implied WarrantiesExclusion of Damages Effect of Local LawMedicalsales@planar.com Medicalsupport@planar.com