VXI 82-0102-000, CT-100C Series Six-Slot VXIBus Chassis user manual FAN Speed Switch Location

Page 24

VXI Technology, Inc.

FAN SPEED SWITCH

A fan speed selector switch is located at the rear of the CT-100C chassis. To set the fans to operate at HIGH speed, move the switch at the rear of the chassis to the left. At HIGH speed, the chassis is provided maximum cooling to the instruments.

The fans will operate in the LOW speed mode when the fan speed switch is toggled to the right. The low speed mode provides for quieter operation, but decreases the cooling capability of the fans. To best determine which mode of operation is best in a given application, refer to Figure 1-6on page 19 that delineates the cooling requirements of populated mainframes. If all instruments installed in the chassis receive adequate cooling in the low speed mode, then this mode may be used.

VXI Technology

WARNING - TO PREVENT POSSIBLE ELECTRIC SHOCK HAZARD, DISCONNECT POWER ! CORD BEFORE REMOVING THE POWER SUPPLY MODULE FROM THE MAINFRAME.

ATTENTION - POUR EMPÊCHER LE RISQUE POSSIBLE DE DÉCHARGE ÉLECTRIQUE, DEBRANCHER LE CORDON DE SECTEUR AVANT D'ENLEVER LE MODULE D'ALIMENTATION D' ÉNERGIE DE L'UNITÉ CENTRALE.

WARNING - FOR PROTECTION FROM ELECTRIC SHOCK, THE GROUNDING CONNECTION ! IN THE POWER CORD MUST BE PROPERLY CONNECTED.

ATTENTION - POUR LA PROTECTION CONTRE LE CHOC ÉLECTRIQUE , LA PRISE DE TERRE AU SOL DANS LE CORDON DE SECTEUR DOIT ÊTRE CORRECTMENT RELIÉE.

WARNING - NO OPERATOR SERVICEABLE PARTS INSIDE. REFER SERVICING TO ! TRAINED SERVICE PERSONNEL.

ATTENTION - AUCUNES PI ÈCES REPARABLES PAR UN OP ÉRATEUR À L'INTÉRIEUR. RÉFÉREZ L'ENTRETIEN AU PERSONNEL QUALIFI É.

FAN SPEED

J200

HIGH / LOW

J201

MONITOR / CONTROL

100-240 VAC

10A MAX.

47 - 63 Hz

100-240 VAC

10A MAX.

47 - 63 Hz

Fan Speed Switch Location

(Rear Panel of CT-100C Mainframe)

FIGURE 1-3: FAN SPEED SWITCH LOCATION

24

CT-100C Installation

Image 24
Contents CT-100C Series VXI Technology, Inc Table of Contents Warranty Limitation of WarrantyCertification Restricted Rights LegendC L a R a T I O N O F C O N F O R M I T Y EMCGeneral Safety Instructions Terms and SymbolsAvoid Electric Shock Support Resources CT-100C Preface Page Introduction IntroductionGeneral Description BackplanePerformance Flexibility 2CT-100CSIX-SLOT Chassis Covers OFFCT-100C Mainframe Features Mainframe Feature LocationsVoltage Monitor Leds Front Panel Voltage MonitorsJ201 Connector PIN Assignments J201 Connector PIN DetailCT-100C Specifications MtbfMttr Input Voltage / Freq Fan atH igh Speed Fan at Low S peed 7CT-100C Dimensional Diagram Installation Remote POWER-ON Option Remote POWER-ON Switch Wiring+5 V Standby Wiring Diagram +5 V Standby UsageFAN Speed Switch FAN Speed Switch LocationRackmount Option Installation OverviewRackmount KIT Installation Parts ListRackmount EAR Installation Diagram Rack Slide Installation 20 Assembly InstructionsRack Slide Installation Diagram Rackmount Door KIT Installation Rack Door Installation Diagram Acrylic Rackmount Door KIT Installation Acrylic Rack Door Installation Diagram Installation of VXI Modules Installation ProcedureDisconnecting the Mainframe VXI Technology, Inc CT-100C Installation Service Information Replicable Parts ListCT-100C Power Supply Replacement Power Supply Replacement Diagram Cleaning the Mainframe Index