VXI 82-0102-000 user manual Rack Slide Installation 20, Assembly Instructions

Page 28

VXI Technology, Inc.

RACK SLIDE INSTALLATION (20” AND 24”)

The 20” and 24” slide hardware kits (Options 63 and 64, respectively) provide standard flanges to mount the slides into standard EIA relay racks. It should be noted that the slide flanges in the front should be installed behind the rack’s front panel mounting rails. This will allow the rackmounting ears to sit flush with the front of the rack. It should also be noted that although the hardware provided in this kit will provide all the necessary components to successfully install the CT-100C into a rack, there are many variations in how EIA relay racks are designed and additional adapter hardware may be required to install the chassis. Please refer to the relay rack manufacture’s catalog for additional options.

This procedure provides instructions for installing the 20” or 24” slide kits. Note that the 20” and 24” slide kits are identical except for the length of the slide units. A 20” slide kit is used when the chassis is flush mounted and a 24” slide kit is used when the chassis is recess mounted.

Required Tools

1.#2 Phillips Screwdriver

2.1/8" Flat Blade Screwdriver

Parts List

QTY

ITEM

VTI P/N

2

Slide, Rack Mount, Steel, 20”

37-0054-020

 

- or -

 

2

Slide, Rack Mount, Steel, 24”

37-0054-024

1

Hardware Kit, Rack Mount Slide, Steel

37-0055-000

6

Screw, 8-32 x 3/8” Pan Head Phillips, Sems Zinc

37-0073-037

6

Screw, 8-32 x 1/4” Pan Head Phillips, Steel/Zinc

37-0074-025

4

Screw, 8-32 x 3/8”, F/H Undercut Phillips, Zinc

37-0115-037

2

Bracket, Slide, Front

41-0108-000

1

Bracket, Slide, Top Side

41-0131-000

1

Bracket, Slide, Bottom

41-0132-000

Assembly Instructions

1.Lay the chassis on a protected work surface on its long side with the voltage monitor LEDs of the chassis facing front with the power switch toward the top.

2.If the rackmount ears were installed flush, remove the rackmount ears and install the front slide bracket to them as shown in the assembly drawing using a flat head screw driver. If the rackmount ears were installed recessed, simply install the slide brackets onto the ears as shown on the following page.

3.Reinstall the rackmount ears (if necessary).

4.Install the top and bottom slide brackets using 1/4” Phillips pan head screws. Note that the slide mounting holes should be oriented below the centerline of the chassis.

28

CT-100C Installation

Image 28
Contents CT-100C Series VXI Technology, Inc Table of Contents Warranty Limitation of WarrantyCertification Restricted Rights LegendC L a R a T I O N O F C O N F O R M I T Y EMCGeneral Safety Instructions Terms and SymbolsAvoid Electric Shock Support Resources CT-100C Preface Page Introduction IntroductionBackplane General DescriptionPerformance Flexibility 2CT-100CSIX-SLOT Chassis Covers OFFCT-100C Mainframe Features Mainframe Feature LocationsVoltage Monitor Leds Front Panel Voltage MonitorsJ201 Connector PIN Assignments J201 Connector PIN DetailMtbf CT-100C SpecificationsMttr Input Voltage / Freq Fan atH igh Speed Fan at Low S peed 7CT-100C Dimensional Diagram Installation Remote POWER-ON Option Remote POWER-ON Switch Wiring+5 V Standby Wiring Diagram +5 V Standby UsageFAN Speed Switch FAN Speed Switch LocationRackmount Option Installation OverviewRackmount KIT Installation Parts ListRackmount EAR Installation Diagram Rack Slide Installation 20 Assembly InstructionsRack Slide Installation Diagram Rackmount Door KIT Installation Rack Door Installation Diagram Acrylic Rackmount Door KIT Installation Acrylic Rack Door Installation Diagram Installation of VXI Modules Installation ProcedureDisconnecting the Mainframe VXI Technology, Inc CT-100C Installation Service Information Replicable Parts ListCT-100C Power Supply Replacement Power Supply Replacement Diagram Cleaning the Mainframe Index