Omega Speaker Systems Calibration Procedure for HX93AC all styles, Recommended equipment

Page 11

HX93A, HX93DA SERIES

11

RH/Temperature Transmitter

11. Calibration

Your transmitter has been factory calibrated to meet or exceed the specifications outlined in this manual. To maintain original specifications it is generally recommended that your transmitter be re- calibrated on an annual basis depending on operating conditions.

12.Calibration Procedure for HX93AC (all styles)

Relative Humidity Adjustment

Recommended equipment:

Humidity Calibration Kit, OMEGA® Model No.: HX92-CAL

Handheld Digital Multi-meter, OMEGA® Model No.: HHM29

Dc Power Supply, OMEGA® Model No.: PSU-93

1.Remove enclosure cover

2.Connect transmitter as shown in figure below.

3.Apply power to transmitter and allow to warm up for 15 min.

4.Place unit in a 11% RH environment and allow to stabilize for 15 min.

5.Adjust potentiometer “P1” for a output reading of 5.76 mA

6.Place unit in a 75% RH environment and allow to stabilize for 15 min.

7.Adjust potentiometer “P2” for a output reading of 16.0 mA

8.Repeat steps 4, 5, 6, and 7 as necessary until proper readings are maintained.

9. Calibration complete.

NOTE:

Do not adjust

P3. This is a

factory

23

1 2

setting.

1

3

TB1

P1 P3

TB2

 

 

P2

HHM29

16.00 mA

SENSOR HEAD

10A uAmA COM V

+ 24V –

PSU-93

HX92-CA

CALIBRATION BOTTLE

Figure 6 - HX93AC Calibration Procedure

7

Relative Humidity Adjustment

Image 11
Contents User’sGuide Servicing Europe Servicing North AmericaFor immediate technical or application assistance Section HX93A, HX93DA SeriesHX93A, HX93DA Series RM/Temperature TransmitterAdditional Transmitter Models Available Description RH/Temperature TransmitterGeneral Description UnpackingFor HX93A For HX93DA Example When using a 24 Vdc power supplyTheory of Operation Complimentary Instruments Recommended AccessoriesMounting Electrical Connections Current Transmitter Wiring ExamplesRelative Humidity Output Calculations Voltage Transmitter Wiring ExamplesVdc ÷ .005848 4 = 81.5ºF Temperature Output CalculationsRH Measured Vs Output Reading For Current output in ºC -20 to 75ºCRecommended equipment Calibration Procedure for HX93AC all stylesRelative Humidity Adjustment CalibrationHX93AC Calibration Procedure Temperature Adjustment Temperature AdjustmentHX93AV Calibration Procedure Relative Humidity Adjustment Calibration Procedure for HX93AV all stylesHX93AV Calibration Procedure Temperature Adjustment Relative Humidity Display Scaling Recommended equipment 15A. Display Scaling for HX93DAC all stylesFor ºC Display Temperature Display Scaling Recommended equipmentFor ºF Display Relative Humidity MaintenanceSpecifications TemperatureSensor Type Ohm Platinum RT General SpecificationsDimensions Weight Wall Mount Model Duct Mount ModelWARRANTY/DISCLAIMER Temperature