HX93A, HX93DA SERIES | 14 |
RH/Temperature Transmitter |
14.Calibration Procedure for HX93AV (all styles)
Relative Humidity Adjustment
Recommended equipment:
•Humidity Calibration Kit, OMEGA® Model No.:
•Handheld Digital
•Dc Power Supply, OMEGA® Model No.:
1.Remove enclosure cover.
2.Connect transmitter as shown in figure below.
3.Apply power to transmitter and allow to warm up for 15 min.
4.Place unit in a 11% RH environment and allow to stabilize for 15 min.
5.Adjust potentiometer “P3” for a output reading of .110 Vdc.
6.Place unit in a 75% RH environment and allow to stabilize for 15 min.
7.Adjust potentiometer “P2” for a output reading of .750 Vdc.
8.Repeat steps 4, 5, 6, and 7 as necessary until proper readings are maintained.
9.Calibration complete.
TB1
SENSOR HEAD
1 2 3
P1 P3
TB2 1 2 3
P2
HHM29
.750 V
10A uAmA COM V
+24V –
Figure 8 - HX93AV Calibration Procedure
Relative Humidity Adjustment
9