HX93A, HX93DA SERIES | 15A | 
RH/Temperature Transmitter | 
15A. Display Scaling for HX93DAC (all styles)
Your HX93DA transmitter comes with dual 
Relative Humidity Display Scaling
Recommended equipment:
•Humidity Calibration Kit, OMEGA® Model No.: 
•Handheld Digital 
•Dc Power Supply, OMEGA® Model No.: 
1.Remove enclosure cover
2.Connect transmitter as shown in figure 6.
3.Apply power to transmitter and allow to warm up for 15 min.
4.Place unit in a 11% RH environment and allow to stabilize for 15 min.
5.Adjust potentiometer "ZERO" on the back of the RH display for a reading of 11.0%
6.Place unit in a 75% RH environment and allow to stabilize for 15 min.
7.Adjust potentiometer "GAIN" on the back of the RH display for a reading of 75.0%
8.Repeat steps 4, 5, 6, and 7 as necessary until proper readings are maintained.
9.
TRANSMITTER LID REAR VIEW
ZERO POTENTIOMETER  | Z  | 
  | 
  | RELATIVE HUMIDITY  | |
  | G  | 
  | 
  | ||
GAIN POTENTIOMETER  | +  | –  | INDICATOR  | ||
  | |||||
ZERO POTENTIOMETER  | Z  | 
  | 
  | TEMPERATURE  | |
  | 
  | 
  | |||
  | 
  | 
  | 
  | ||
GAIN POTENTIOMETER  | G  | +  | –  | INDICATOR  | |
  | 
  | ||||
  | 
  | 
  | 
Figure 10 - Transmitter Lid Rear View
11