15 | HX93A, HX93DA SERIES |
RH/Temperature Transmitter | |
|
|
15.Calibration Procedure for HX93AV (all Styles)
Temperature Adjustment
Recommended equipment:
•100 Ohm RTD Simulator OMEGA® Model No.:
•Handheld Digital
•Dc Power Supply, OMEGA® Model No.:
1.Remove enclosure cover
2.Connect the power supply and multimeter as shown in the figure below.
3.Disconnect the
4.Connect the RTD simulator to the “J4” header on the main board as shown below.
5.Apply power to transmitter and allow to warm up for 10 min.
6.Set the RTD simulator to the 0ºF position.
7.Adjust potentiometer “P5” on the main board for a output reading of 0.023 Vdc
8.Set the RTD simulator to the 150ºF position.
9.Adjust potentiometer “P4” for a output reading of 0.901 Vdc
10.Repeat steps 6, 7, 8 and 9 as necessary until proper readings are maintained.
11. Calibration complete.
1 2 3 | 1 23 |
TB1 | TB2 |
P5
P4
J4
SENSOR HEAD
HHM29
.901 V
10A uAmA COM V
+24V –
Figure 9 - HX93AV Calibration Procedure
Temperature Adjustment
10