Select Low-point Calibration
This is the Calibration Status display. This display shows that no input is selected for calibration.
∙ nonE: | No selection |
∙ip1.L: Input 1 (PV1) calibration
∙ip1.H: Input 1 (PV1) calibration
∙ip2.L: Input 2 (PV2) calibration
∙ip2.H: Input 2 (PV2) calibration
Use / to select the parameter for the Low Calibration point of Input 1, ‘ip1.L’.
Press
Adjust low-point calibration
This is the display for adjusting the Low Calibration point of Input 1. The lower readout is a live reading of the process value, which changes as the input changes.
Make sure that the calibration source is connected to the terminals of Input 1, switched on and feeding a signal to the controller. It should be set to the desired
value, then use / to adjust the reading to the required value.
Press to return to the
To perform the
adjustment.
Press three times.
Calibration type
‘USEr’ was selected for the
Press