User calibration
With fixed reference points
0 point / 1/3 load / Max. load
Switch the scale off. Switch the scale on.
Press the
and
keys and hold for 2 sec. during
Empty the scale and confirm the zero point by pressing
.
Place the default weight on the scale (1/3 load) and con- firm by pressing
..
Place the default weight on the scale (Max. load) and confirm by pressing
.
Empty the scale.
The scale goes to weighing mode.
Nullstellen
With variable reference point 0 point Test weight < Max
Switch the scale off. Switch the scale on.
Press the
and
keys and hold for 2 sec. when switching on.
Empty the scale and confirm the zero point by pressing
the
key.
You can toggle between the decades by pressing the ![]()
and
keys.
You can increment the numerical value by pressing the
key.
Enter the existing weight, e.g. 6kg, and confirm by pres- sing
.
Place the existing weight on the scale and confirm by pressing
.
Empty the scale.
The scale goes to weighing mode.