Agilent Technologies 6010a, 6011A, 6012B Connecting a Bypass Capacitor Overvoltage Protection OVP

Page 30

Either positive or negative voltages can be obtained from the supply by grounding one of the output terminals. It is best to avoid grounding the output at any point other than the power supply output terminals to avoid noise problems caused by common-mode current flowing through the load leads to ground. Always use two wires to connect the load to the supply regardless of where or how the system is grounded. Never ground the system at more than one point. The maximum potential (including output voltage) that either output terminal is from ground must not exceed that specified on the output label on the rear chassis.

The PARD specifications in Table 1-1 apply at the power supply output terminals. However, noise spikes induced in the load leads at or near the load may affect the load although the spikes are inductively isolated from the power supply. To minimize voltage spikes at the load, connect a bypass capacitor as shown in Figure 3-2. With this setup, peak-to-peak noise at the load can actually be reduced to a level below the value specified at the power supply output terminals.

500

Figure 3-2. Connecting a Bypass Capacitor

Overvoltage Protection (OVP)

The overvoltage trip point is adjusted at the front panel. The approximate trip voltage range is from zero volts to approximately 107% of maximum rated voltage of the power supply. When the OVP circuit trips, the power supply output is disabled and delivers no output power, and the OVP and UNREGULATED indicators turn on.

Adjustment. OVP is set by the recessed single-turn OVP ADJUST potentiometer on the front panel. Rotating the control clockwise sets the trip voltage higher. (It is set to maximum at the factory.) When adjusting the OVP trip point, the possibility of false tripping must be considered. If the trip voltage is set too close to the supply’s operating voltage, a transient in the output would falsely trip the OVP. For this reason it is recommended that the OVP trip voltage be set higher than the output voltage by at least 1 volt. To adjust the OVP trip voltage, proceed as follows:

a.Turn on supply and hold DISPLAY OVP pushbutton in.

b.Insert a small-blade screwdriver through hole in front panel and adjust OVP trip voltage to desired level.

OVP Reset. To reset OVP, turn the LINE switch off and then back on. The cause of the overvoltage must be removed before the OVP circuit is reset or the circuit will trip again immediately. If the OVP circuit trips continuously check the load and the trip voltage.

Protective Shutdown

Protective circuits within the power supply may limit or turn off the output in case of abnormal conditions. The cause of the protective action can be determined by observing the front panel indicators (lights and meters).

Unregulated. If an overrange condition exists (load tries to draw more power than the supply can deliver), the UNREGULATED indicator turns on and both the CV and CC indicators are off. The product of the VOLTS and AMPS displays will exceed the maximum output power of the supply. Also, if the power supply output is disabled by protective

30 Operating Instructions

Image 30
Contents USER’S Guide Certification Safety Summary Safety Symbols Product Names Safety InformationManufacturer’s Name and Address Model NumbersHerstellerbescheinigung Manufacturer’s DeclarationTable Of Contents VAC Input Power Option IndexIntroduction Safety ConsiderationsGeneral Information DescriptionFor Agilent Models 6023A and 6028A AccessoriesOption Description Agilent Part No DescriptionSpecifications Instrument and Manual IdentificationOrdering Additional Manuals Related Documents 6010A Performance SpecificationsSupplemental Characteristics Agilent Technologies Model 6011A6023A 6012B6015A 6028AShipping Supplemental CharacteristicsTemperature Rating C CertificationOutput Characteristic Curve General Information Output Impedance General Information Installation Initial InspectionPreparation for Use Outline Diagrams Input Power RequirementsRack Mounting Power Connection Agilent Models 6010A, 6011A, 6012B and 6015AAgilent Models 6023A and 6028A Line Voltage Option Conversion Agilent 6010A, 6011A, 6012B Line Voltage Conversion Components Installation AC Line Impedance Check Repackaging for ShipmentModel Voltage Rear Panel Screw Sizes and Part Numbers Description Agilent Part numberAgilent Model 6023A and 6033A Front-Panel Controls and Indicators Operating Instructions Operating InstructionsTurn-On Checkout Procedure Display OVPOperating Instructions Initial Setup and Interconnections Connecting the LoadMaximum Wire Lengths To Limit Voltage Drops Stranded Copper Wire AmpacityWire Size Connecting a Bypass Capacitor Overvoltage Protection OVP Protective ShutdownNormal Mode Operating ModesDetermining Operating Point Overrange Operation Constant Voltage Operation Remote Voltage Sensing Constant Current OperationVoltage change due to open sense lead BothAnalog Programming Constant Voltage Output, Resistance ControlVoltage Programming of Output Voltage Operating Instructions = Handle Multiple-Supply Operation Constant Current Output, Voltage ControlAuto-Parallel Operation 12. Auto-Parallel Operation Series OperationMonitor Signals 13. Series OperationPage General Information VAC Input Power OptionUsing Appendix a Manual ChangesManual Changes Index Index Europe Japan Agilent Sales and Support OfficeUnited States Canada Latin America Australia/New ZealandManual Updates
Related manuals
Manual 126 pages 62.23 Kb

6015A, 6023A, 6028A, 6012B, 6011A, 6010a specifications

Agilent Technologies, a leader in the field of measurement and analysis, offers a suite of instruments within its 6010 and 6011 series, specifically the 6010A, 6011A, 6012B, and 6015A models. These devices are designed to meet the needs of various industries, including healthcare, environmental monitoring, and materials testing.

The Agilent 6010A is a high-performance spectrometer known for its precision and versatility. It utilizes advanced optical technologies to provide exceptional wavelength accuracy and resolution. This model is particularly useful in laboratories where reliable data is critical, offering a wide spectral range and effective noise reduction features. Its user-friendly interface simplifies complex analyses, making it suitable for both seasoned professionals and newcomers.

Following closely, the Agilent 6011A is recognized for its robust capabilities in laboratory environments. This device incorporates advanced signal processing techniques, enabling high-throughput measurements without compromising on quality. The 6011A is ideal for real-time monitoring applications, ensuring that users can make informed decisions based on accurate, timely data. Its comprehensive software suite is designed to enhance data analysis, allowing for seamless integration with existing laboratory workflows.

The 6012B variant enhances the functionality further by introducing additional features tailored for specific applications. With a focus on flexibility, the 6012B supports multiple measurement modes, including direct and differential detection. This model excels in complex measurements, allowing for greater analytical depth and insights. The built-in calibration options ensure consistent performance, making it a reliable choice for various research and development tasks.

Lastly, the Agilent 6015A model stands out with its leading-edge technology, designed for the most demanding applications. It boasts enhanced sensitivity and an improved dynamic range, making it perfect for trace analysis in challenging environmental samples. The 6015A’s advanced reporting tools provide detailed analytics, helping scientists and researchers interpret results efficiently. Its compact design also makes it suitable for laboratory spaces with limited room, without sacrificing performance.

Together, these models showcase Agilent Technologies' commitment to delivering high-quality, innovative solutions that empower users to achieve their analytical goals effectively and efficiently. Whether in a research, clinical, or industrial setting, the 6010A, 6011A, 6012B, and 6015A continue to set standards in precision instrumentation.