Agilent Technologies 6023A, 6028A, 6010a, 6011A AC Line Impedance Check, Repackaging for Shipment

Page 23

AC Line Impedance Check

The power supply is designed for proper operation with line impedances typically found in ac power lines. However, if the supply is connected to an ac power line having a high impedance combined with line voltage near the minimum specified value, (e.g., 104 Vac for nominal 120 Vac), the unit will go out of regulation if it is asked to provide full rated output power. Such a situation might occur if the supply is connected to ac power an extended distance from the main ac distribution terminals and/or if the ac power wires from the main ac distribution terminals are of relatively small gauge.

Measurement of ac line voltage at the supply input terminals typically is not a reliable indication of the actual ac line voltage because of the peak clipping effect of the power supply and the averaging effect of the voltmeter. Symptoms of excessive line impedance may include erratic or no output from the supply and/or inability of the supply to provide full output power. If there is reason to suspect the ac power lines to the supply may have high impedance, perform the following check:

WARNING: This check should be performed only by service-trained personnel who are aware of the hazards involved (for example, fire and electrical shock). Turn the power supply off before making or breaking any connections to the power supply. Hazardous voltages are present within the unit even when the power switch is turned off.

a.Connect a variable load to the supply. Using the OUTPUT ADJUST controls and DISPLAY SETTINGS, set voltage and current (see Section III for detailed description) to maximum rating.

b.Set the load to the maximum rated output current for the power supply (see Appendix A, Figure A-1). The power supply output voltage should be greater than:

Model

Voltage

Agilent 6010A

65V

Agilent 6011A

8V

Agilent 6612B

22V

Agilent 6015A

220V

Agilent 6023A

6V

Agilent 6028A

20V

c.If the supply voltage is less than specified, perform the power limit calibration given in the Service Manual. If the power limit is calibrated correctly, but the unit still does not provide the required output, then the power supply is not receiving adequate ac line input.

Repackaging for Shipment

To insure safe shipment of the instrument, it is recommended that the package designed for the instrument be used. The original packaging material is reusable. If it is not available, contact your local Agilent Technologies Sales and Support office to obtain the materials. This office will also furnish the address of the nearest service office to which the instrument can be shipped. Be sure to attach a tag to the instrument specifying the owner, model number, full serial number, and service required or a brief description of the trouble.

Installation 23

Image 23
Contents USER’S Guide Certification Safety Summary Safety Symbols Model Numbers Safety InformationManufacturer’s Name and Address Product NamesManufacturer’s Declaration HerstellerbescheinigungTable Of Contents Index VAC Input Power OptionDescription Safety ConsiderationsGeneral Information IntroductionAgilent Part No Description AccessoriesOption Description For Agilent Models 6023A and 6028AOrdering Additional Manuals Related Documents SpecificationsInstrument and Manual Identification 6011A Performance SpecificationsSupplemental Characteristics Agilent Technologies Model 6010A6028A 6012B6015A 6023ACertification Supplemental CharacteristicsTemperature Rating C ShippingOutput Characteristic Curve General Information Output Impedance General Information Preparation for Use InstallationInitial Inspection Rack Mounting Outline DiagramsInput Power Requirements Agilent Models 6023A and 6028A Power ConnectionAgilent Models 6010A, 6011A, 6012B and 6015A Line Voltage Option Conversion Agilent 6010A, 6011A, 6012B Line Voltage Conversion Components Installation Model Voltage AC Line Impedance CheckRepackaging for Shipment Agilent Model 6023A and 6033A Rear Panel Screw Sizes and Part NumbersDescription Agilent Part number Operating Instructions Front-Panel Controls and Indicators Operating InstructionsDisplay OVP Turn-On Checkout ProcedureOperating Instructions Connecting the Load Initial Setup and InterconnectionsWire Size Maximum Wire Lengths To Limit Voltage DropsStranded Copper Wire Ampacity Protective Shutdown Connecting a Bypass Capacitor Overvoltage Protection OVPOperating Modes Normal ModeDetermining Operating Point Overrange Operation Constant Voltage Operation Constant Current Operation Remote Voltage SensingBoth Voltage change due to open sense leadConstant Voltage Output, Resistance Control Analog ProgrammingVoltage Programming of Output Voltage Operating Instructions = Handle Auto-Parallel Operation Multiple-Supply OperationConstant Current Output, Voltage Control Series Operation 12. Auto-Parallel Operation13. Series Operation Monitor SignalsPage Manual Changes VAC Input Power OptionUsing Appendix a General InformationManual Changes Index Index Latin America Australia/New Zealand Agilent Sales and Support OfficeUnited States Canada Europe JapanManual Updates
Related manuals
Manual 126 pages 62.23 Kb

6015A, 6023A, 6028A, 6012B, 6011A, 6010a specifications

Agilent Technologies, a leader in the field of measurement and analysis, offers a suite of instruments within its 6010 and 6011 series, specifically the 6010A, 6011A, 6012B, and 6015A models. These devices are designed to meet the needs of various industries, including healthcare, environmental monitoring, and materials testing.

The Agilent 6010A is a high-performance spectrometer known for its precision and versatility. It utilizes advanced optical technologies to provide exceptional wavelength accuracy and resolution. This model is particularly useful in laboratories where reliable data is critical, offering a wide spectral range and effective noise reduction features. Its user-friendly interface simplifies complex analyses, making it suitable for both seasoned professionals and newcomers.

Following closely, the Agilent 6011A is recognized for its robust capabilities in laboratory environments. This device incorporates advanced signal processing techniques, enabling high-throughput measurements without compromising on quality. The 6011A is ideal for real-time monitoring applications, ensuring that users can make informed decisions based on accurate, timely data. Its comprehensive software suite is designed to enhance data analysis, allowing for seamless integration with existing laboratory workflows.

The 6012B variant enhances the functionality further by introducing additional features tailored for specific applications. With a focus on flexibility, the 6012B supports multiple measurement modes, including direct and differential detection. This model excels in complex measurements, allowing for greater analytical depth and insights. The built-in calibration options ensure consistent performance, making it a reliable choice for various research and development tasks.

Lastly, the Agilent 6015A model stands out with its leading-edge technology, designed for the most demanding applications. It boasts enhanced sensitivity and an improved dynamic range, making it perfect for trace analysis in challenging environmental samples. The 6015A’s advanced reporting tools provide detailed analytics, helping scientists and researchers interpret results efficiently. Its compact design also makes it suitable for laboratory spaces with limited room, without sacrificing performance.

Together, these models showcase Agilent Technologies' commitment to delivering high-quality, innovative solutions that empower users to achieve their analytical goals effectively and efficiently. Whether in a research, clinical, or industrial setting, the 6010A, 6011A, 6012B, and 6015A continue to set standards in precision instrumentation.