Agilent Technologies 6015A, 6010a Supplemental Characteristics, Temperature Rating C, Shipping

Page 14

Table 1-2 Supplemental Characteristics (continued)

DC Floating Voltage: Either output terminal may be floated up to the following voltage (including the output voltage) from earth ground:

±240 Vdc on Models 6011A, 6012B, 6023A, and 6028A

±550 Vdc on Models 6010A and 6015A

Exceeding these voltages can result in damage to the equipment.

Efficiency (typical): 80% on maximum output boundary

Remote Sensing: The power supply maintains specifications at the load with up to 0.5 Volt per load lead with sense wire resistance less than 0.2 Ω per lead and sense lead length less than 5 metres. Operation with up to 2 volts per load lead is possible with some degradation of the load effect specification.

Multiple Operations: Up to two similar units may be connected in series, parallel or auto-parallel, to provide increased output capabilities. Mixing supplies with dissimilar output capabilities is not recommended because under certain conditions, the lower output supply may be stressed beyond its maximum voltage and or current capabilities by the higher output supply.

Reactive Loads: Stable with inductive loads up to 100 mH and capacitive loads up to 10 F. CC compensation that provides up to 10 H (with increased settling times) is available on special order.

Voltage Overshoot (typical): The output voltage will overshoot its steady state value by less than 250 mV (1 V on Model

6015A) due to any of the following conditions:

1.Up programming

2.Crossover from CC to CV mode

3.A step change of up to 5A

4.AC power on

Temperature Rating (°C):

Operating is 0-50 (Agilent 6010A/6011A/6012B/6015A) 0-55 (Agilent 6023A/6028A)

Storage is - 40 + 75 (all models)

Weight kg. (Lbs.)

 

 

 

 

 

Model

Agilent 6010A

Agilent 6011A

Agilent6012B

Agilent 6015A

Agilent 6023A

Agilent 6028A

Net

15.9 (35)

16.8 (37)

15.9 (35)

16.3 (36)

8.6 (19)

8.6 (19)

Shipping

21.3 (47)

22.3 (49)

21.4 (47)

21.7 (48)

10.5 (23)

10.5 (23)

Dimensions: See Figure 2-1.

Certification:

The unit is designed to comply with these requirements:

IEC 348-Safety Requirements for Electronic Measuring Apparatus.

CSA Electrical Bulletin 556B-Electronic Instruments and Scientific Apparatus for Special Use and Applications.

VDE 0871.6.78 Level B-RFI Suppression of Radio Frequency Equipment for Industrial, Scientific, and Medical (ISM) and similar purposes.

VDE 0411-Electronic Measuring Instruments and Automatic Controls.

UL 1244-Electrical and Electronic Measuring & Testing Equipment.

ANSI C39.5 Part 0 Draft 8-Electrical Testing, Measurement, and Control Equipment.

Agilent Class B – Environmental Specifications

14 General Information

Image 14
Contents USER’S Guide Certification Safety Summary Safety Symbols Product Names Safety InformationManufacturer’s Name and Address Model NumbersHerstellerbescheinigung Manufacturer’s DeclarationTable Of Contents VAC Input Power Option IndexIntroduction Safety ConsiderationsGeneral Information DescriptionFor Agilent Models 6023A and 6028A AccessoriesOption Description Agilent Part No DescriptionOrdering Additional Manuals Related Documents SpecificationsInstrument and Manual Identification 6010A Performance SpecificationsSupplemental Characteristics Agilent Technologies Model 6011A6023A 6012B6015A 6028AShipping Supplemental CharacteristicsTemperature Rating C CertificationOutput Characteristic Curve General Information Output Impedance General Information Preparation for Use InstallationInitial Inspection Rack Mounting Outline DiagramsInput Power Requirements Agilent Models 6023A and 6028A Power ConnectionAgilent Models 6010A, 6011A, 6012B and 6015A Line Voltage Option Conversion Agilent 6010A, 6011A, 6012B Line Voltage Conversion Components Installation Model Voltage AC Line Impedance CheckRepackaging for Shipment Agilent Model 6023A and 6033A Rear Panel Screw Sizes and Part NumbersDescription Agilent Part number Front-Panel Controls and Indicators Operating Instructions Operating InstructionsTurn-On Checkout Procedure Display OVPOperating Instructions Initial Setup and Interconnections Connecting the LoadWire Size Maximum Wire Lengths To Limit Voltage DropsStranded Copper Wire Ampacity Connecting a Bypass Capacitor Overvoltage Protection OVP Protective ShutdownNormal Mode Operating ModesDetermining Operating Point Overrange Operation Constant Voltage Operation Remote Voltage Sensing Constant Current OperationVoltage change due to open sense lead BothAnalog Programming Constant Voltage Output, Resistance ControlVoltage Programming of Output Voltage Operating Instructions = Handle Auto-Parallel Operation Multiple-Supply OperationConstant Current Output, Voltage Control 12. Auto-Parallel Operation Series OperationMonitor Signals 13. Series OperationPage General Information VAC Input Power OptionUsing Appendix a Manual ChangesManual Changes Index Index Europe Japan Agilent Sales and Support OfficeUnited States Canada Latin America Australia/New ZealandManual Updates
Related manuals
Manual 126 pages 62.23 Kb

6015A, 6023A, 6028A, 6012B, 6011A, 6010a specifications

Agilent Technologies, a leader in the field of measurement and analysis, offers a suite of instruments within its 6010 and 6011 series, specifically the 6010A, 6011A, 6012B, and 6015A models. These devices are designed to meet the needs of various industries, including healthcare, environmental monitoring, and materials testing.

The Agilent 6010A is a high-performance spectrometer known for its precision and versatility. It utilizes advanced optical technologies to provide exceptional wavelength accuracy and resolution. This model is particularly useful in laboratories where reliable data is critical, offering a wide spectral range and effective noise reduction features. Its user-friendly interface simplifies complex analyses, making it suitable for both seasoned professionals and newcomers.

Following closely, the Agilent 6011A is recognized for its robust capabilities in laboratory environments. This device incorporates advanced signal processing techniques, enabling high-throughput measurements without compromising on quality. The 6011A is ideal for real-time monitoring applications, ensuring that users can make informed decisions based on accurate, timely data. Its comprehensive software suite is designed to enhance data analysis, allowing for seamless integration with existing laboratory workflows.

The 6012B variant enhances the functionality further by introducing additional features tailored for specific applications. With a focus on flexibility, the 6012B supports multiple measurement modes, including direct and differential detection. This model excels in complex measurements, allowing for greater analytical depth and insights. The built-in calibration options ensure consistent performance, making it a reliable choice for various research and development tasks.

Lastly, the Agilent 6015A model stands out with its leading-edge technology, designed for the most demanding applications. It boasts enhanced sensitivity and an improved dynamic range, making it perfect for trace analysis in challenging environmental samples. The 6015A’s advanced reporting tools provide detailed analytics, helping scientists and researchers interpret results efficiently. Its compact design also makes it suitable for laboratory spaces with limited room, without sacrificing performance.

Together, these models showcase Agilent Technologies' commitment to delivering high-quality, innovative solutions that empower users to achieve their analytical goals effectively and efficiently. Whether in a research, clinical, or industrial setting, the 6010A, 6011A, 6012B, and 6015A continue to set standards in precision instrumentation.