Siemens 1015N-2MFM-1A manual Setting 1010N Flowmeter RS-232 Parameters, Telephone Line Connections

Page 12

1015N-2MFM-1A

8.Install the 1015N-2M Modem Card (7) on top of the 1010N-8M module (2) carefully engaging connector P1 on the 1015N-2M with its mate P2 on the 1010N-8M module. Secure the 1015N-2M by engaging its four captive screws (1) with the four standoffs (4) on the 1010N- 8M module.

9.Install the large threaded cover mount standoff (8) from the mounting kit through the hole at the center of the 1015N-2M Modem Card, through the standoff on the 1010N-8M module and then into the threaded receptacle on the 1010N-2M module shield (9).

10.Referring to installation drawing 1015N-2M-7 (see Appendix A), plug a telephone jack into J4 of the Modem Card. Thread the telephone line plug through one of the System 1010 flowmeter access holes and plug it into a standard telephone outlet (refer to paragraph

5.3below for communication connection options).

11.Replace Access Cover (10) and finger tighten captive thumbscrew (11).

12.Place the 1010 flowmeter power switch to the ON position.

13.Close the 1010 flowmeter top cover by locking the cover latch.

14.Proceed to paragraph 5.2 to set 1010N flowmeter RS-232 parameters.

5.2SETTING 1010N FLOWMETER RS-232 PARAMETERS

Set the RS-232 parameters on the 1010N flowmeter to match the 1015N-2M Modem Card as follows:

lBaud Rate: 2400

lParity: Odd

lData Bits: 7

lLine Feed: Yes

5.3TELEPHONE LINE CONNECTIONS

The 1015N-2M Modem Card contains an FCC-approved Data Access Arrangement (DAA) and may be connected to your conventional subscriber telephone line. It is suggested that you consult with your phone service provider and/or telecommunications specialist when connecting the modem to other types of telephone networks.

If the equipment is installed in a location where lightning strikes are possible, reduce this risk by providing your phone line and AC-power input with surge and lightning protection.

Connection to the telephone line is made in one of the follows ways:

1.Directly plug your telephone line into the modular connector RJ11 on PE2400 of the Modem Card (refer to Figure 1 and installation drawing 1015N-2M-7), or

2.Connect the two modular connectors together on the Modem Card with the supplied Modem Cable Assembly (P/N 1015-154) and then plug your telephone line to the 2-wire terminal block TB1 on the Modem Card (refer to drawing 1015N-2M-7 in Appendix B).

NOTE: When connecting your telephone line observe the following:

Connect the Green wire (Ring Signal) to TB1-1 and the Red wire (TIP) to TB1-2.

Once the telephone line connection is completed, simply turn the 1010N flowmeter power on and the modem is ready for use. The modem’s communications parameters are re-loaded every time power is applied.

8

Image 12
Contents Field Manual Important Notice Manual Changes Table of Contents Introduction 1015N-2M Modem Card Kits Option 1 Mounting KIT 1015N-2M-MK1Option 2 Mounting KIT 1015N-2M-MK2 Option 3 Mounting KIT 1015N-2M-MK3Technical Specifications Preliminary Installation Notes and Cautions1010N Flowmeter System Requirements 1015N-2M Modem Card Installation Option 1 Installation side view Refer to Installation Drawing 1015N-2M-MK-7 sheet 2 Refer to Installation Drawing 1015N-2M-MK-7 sheet 3 Option 3 Installation side view Setting 1010N Flowmeter RS-232 Parameters Telephone Line ConnectionsCommunications Configuration Windows HyperTerminalTM SetupAT Command SET and S Registers Appendix aAT Commands Command Function Basic AT CommandsSelect MNP 10 link negotiation at highest rate Registers AT Commands and S Registers Dial ModifiersAT Commands and ‘S Registers Ampersand AT& CommandsPercent AT% Commands Atcommands and S Registers Appendix B Engineering Drawings Page Page