Programming Reference

Gating Time

:SENS1:GATE:PER

Syntax :SENS1:GATE:PER? :SENS1:GATE:PER integer

Description Either gets or sets the duration of the gating interval. The allowed range for the gating time is 1s to 608399s (168 hours, 59 minutes, 59 seconds). The gating time is always in seconds. Fractional seconds are not supported, and will be rounded to the closest time in seconds.

Gating Bit Error Rate

:SENS1:FETC:ERAT

Syntax :SENS1:FETC:ERAT?

Description Gets the bit error ration that is captured during the gating interval. This value might/will change while the gating is in progress, and will freeze at the end of the gating interval.

Gating Number of Errors

:SENS1:FETC:ECO

Syntax :SENS1:FETC:ECO?

Description Gets the number of errors that have been counted during the gating interval. This value might/will change while gating is in progress, and will freeze at the end of the gating interval.

Start Gating

:SENS1:GATE:STAT 1

Syntax :SENS1:GATE:STAT 1ON

Description Starts a new gating period. This command will be ignored if a gating period is already in progress.

Abort Gating

:SENS1:GATE:STAT 0

Syntax :SENS1:GATE:STAT 0OFF

Description This command aborts the currently active gating period immediately. It will be ignored if the gating is not in progress.

40

N5980A User Guide

Image 40
Contents Agilent 125 Gb/s Serial Bert Revision 1.0 May Herrenberger StrSafety Summary Environmental Information Contents Programming Reference Introduction Inspect Shipment CD content isSerial Bert Connect the Instrument Front PanelRear Panel To Connect N5980AInstall User Software for N5980A Serial Bert setup diagramTo Install the User Software for N5980A Select the folder in which N5980A will be saved To Access N5980A Demo Offline Advanced Mode IntroductionRunning the BER Test Setting the Instrument Parameters SetupAdvanced Viewing the Results BER displays Gating BER and the Elapsed Gating TimeStore Setting Advanced ModeFile Menu Connection Menu Connect Each PC can control only one N5980AHelp Menu Error IndicatorsInfo BER Bar ErrorSync Loss General Setup and Advanced TabGenerator Error Detector InputSFP Data Pattern SMA Data PatternBER and Advanced BER Tab Gating BER BarGating Errors Actual ErrorsError Detector Pattern Advanced BERCommon Command Summary Agilent N5980A Command SummaryCommon Commands Scpi Instrument Command List- ReferenceID Query Read error queue Load factory default settingOperation complete Option QueryStore current setting Recall settingAll Channels DatarateSMA Output Electrical Generator PatternSOUR1PATT Either sets or gets the pattern of the SMA outputError Insertion Add Single ErrorInsertion Rate Output LevelSOUR2PATT SFP Output Optical GeneratorEither sets or gets the pattern of the SFP output SOUR2PATTEADD Once SOUR2PATTEADDRATESyntax SOUR2PATTEADD Once Trigger Output Detected PatternSOUR3PATT SENS1PATT?Actual Bit Error Rate Synchronization modeActual Number of Errors Gating Bit Error Rate Gating Number of ErrorsGating Time Start GatingDescription Selects the input for the error detector Gating ProgressInput Selection SENS1INPSample Code Communicate with the instrument Programming Reference