Page 17

Operating N5980A with the User Interface

Running the BER Test

The User Software supports a quick pass/fail measurement, which is generally used for testing the DUT in manufacturing. This measurement allows you to count the number of bit errors during a user specified gating time. The result of the measurement is displayed as Gating Errors, Gating BER, and by a simple red (=fail)

/green (=pass) color coding of the Gating BER bar.

1To run the basic measurement, click on Start Gating Period or F12.

2In case you want to stop the running measurement, click on:

3Once the measurement is complete, the results would appear as follows:

NOTE

The Gating Time can be modified in Advanced Mode.

 

 

N5980A User Guide

17

Image 17
Contents Agilent 125 Gb/s Serial Bert Herrenberger Str Revision 1.0 MaySafety Summary Environmental Information Contents Programming Reference Introduction Serial Bert CD content isInspect Shipment Front Panel Connect the InstrumentTo Connect N5980A Rear PanelSerial Bert setup diagram Install User Software for N5980ATo Install the User Software for N5980A Select the folder in which N5980A will be saved To Access N5980A Demo Offline Introduction Advanced ModeRunning the BER Test Setup Setting the Instrument ParametersAdvanced BER displays Gating BER and the Elapsed Gating Time Viewing the ResultsFile Menu Advanced ModeStore Setting Connection Menu Each PC can control only one N5980A ConnectInfo Error IndicatorsHelp Menu Sync Loss ErrorBER Bar Generator Setup and Advanced TabGeneral Input Error DetectorSFP Data Pattern SMA Data PatternGating BER Bar BER and Advanced BER TabActual Errors Gating ErrorsError Detector Pattern Advanced BERAgilent N5980A Command Summary Common Command SummaryID Query Scpi Instrument Command List- ReferenceCommon Commands Load factory default setting Read error queueOperation complete Option QueryRecall setting Store current settingAll Channels DataratePattern SMA Output Electrical GeneratorSOUR1PATT Either sets or gets the pattern of the SMA outputAdd Single Error Error InsertionInsertion Rate Output LevelEither sets or gets the pattern of the SFP output SFP Output Optical GeneratorSOUR2PATT Syntax SOUR2PATTEADD Once SOUR2PATTEADDRATESOUR2PATTEADD Once Detected Pattern Trigger OutputSOUR3PATT SENS1PATT?Actual Number of Errors Synchronization modeActual Bit Error Rate Gating Number of Errors Gating Bit Error RateGating Time Start GatingGating Progress Description Selects the input for the error detectorInput Selection SENS1INPSample Code Communicate with the instrument Programming Reference