Agilent Technologies N5980A manual Gating Progress, Input Selection, SENS1INP

Page 41

Programming Reference

Gating Progress

:SENS1:GATE:STAT

Syntax :SENS1:GATE:STAT?

Description Requests the current state of the gating interval. It will return 1 while the gating is in progress and 0 if the gating is finished/ inactive. It is guaranteed that the very first :SENS1:GATE:STAT? query after the start of the gating period will report a valid state of the gating progress. There is no need to capture 0-1-0 transitions of the gating progress.

Input Selection

SENS1:INP

 

Syntax SENS1:INP

SFPSMA

Description Selects the input for the error detector.

N5980A User Guide

41

Image 41
Contents Agilent 125 Gb/s Serial Bert Herrenberger Str Revision 1.0 MaySafety Summary Environmental Information Contents Programming Reference Introduction Serial Bert CD content isInspect Shipment Front Panel Connect the InstrumentTo Connect N5980A Rear PanelSerial Bert setup diagram Install User Software for N5980ATo Install the User Software for N5980A Select the folder in which N5980A will be saved To Access N5980A Demo Offline Introduction Advanced ModeRunning the BER Test Setup Setting the Instrument ParametersAdvanced BER displays Gating BER and the Elapsed Gating Time Viewing the ResultsFile Menu Advanced ModeStore Setting Connection Menu Each PC can control only one N5980A ConnectInfo Error IndicatorsHelp Menu Sync Loss ErrorBER Bar Generator Setup and Advanced TabGeneral Input Error DetectorSFP Data Pattern SMA Data PatternGating BER Bar BER and Advanced BER TabActual Errors Gating ErrorsError Detector Pattern Advanced BERAgilent N5980A Command Summary Common Command SummaryID Query Scpi Instrument Command List- ReferenceCommon Commands Load factory default setting Read error queueOperation complete Option QueryRecall setting Store current settingAll Channels DataratePattern SMA Output Electrical GeneratorSOUR1PATT Either sets or gets the pattern of the SMA outputAdd Single Error Error InsertionInsertion Rate Output LevelEither sets or gets the pattern of the SFP output SFP Output Optical GeneratorSOUR2PATT Syntax SOUR2PATTEADD Once SOUR2PATTEADDRATESOUR2PATTEADD Once Detected Pattern Trigger OutputSOUR3PATT SENS1PATT?Actual Number of Errors Synchronization modeActual Bit Error Rate Gating Number of Errors Gating Bit Error RateGating Time Start GatingGating Progress Description Selects the input for the error detectorInput Selection SENS1INPSample Code Communicate with the instrument Programming Reference