Teledyne HPM-2002-OBE instruction manual Gas Selection Table, GAS Mode, Analog Output

Page 11

coefficient, place the HPM-2002-OBE in the Cal mode using the SELECT button. Then turn the ADJUST rotary encoder. The Cal light will begin to flash during adjustment and will continue to flash until the Cal mode is exited using the SELECT button.

To perform a full calibration on the HPM-2002-OBE, first use the Zero Coefficient Adjustment Procedure. Followed by the Midrange Coefficient Adjustment Procedure, and finally perform the Atmosphere Coefficient Adjustment Procedure. Sensor Coefficients are stored in the Sensor’s EEPROM upon exiting the Cal mode.

3.1.5.GAS Mode

The HPM-2002-OBE can provide true pressure measurements in many gas environments. At pressure levels where the direct force piezoresistive sensor is operative, the instrument is gas composition independent and measures the true pressure regardless of gas composition. The Pirani is gas composition sensitive so the actual composition must be known and the Pirani calibrated in that gas. When the vacuum system’s gas composition is dominated by a single gas species (for example, during system venting with an inert gas), the user can enter a gas selection into the HPM-2002-OBE by rotating the ADJUST rotary encoder. To view the gas selection, depress the SELECT button until the GAS light is illuminated. The number on the display corresponds to the gas. See the table below.

Gas Selection Table

Gas Mode

Gas

Displayed Number

 

 

 

00

Nitrogen

 

 

01

Argon

 

 

02

Helium

 

 

03

Water Vapor

 

 

04

Custom

 

 

3.1.6.Analog Output (0-10V)

The dual (0 - 10V) output option board provides voltage outputs proportional to the HPM-2002-OBE’s pressure reading. The first channel (pin 5) corresponds to the higher-pressure range (0 - 1024 Torr). The second channel (pin 7) corresponds to the lower pressure range (0 - 1000 mTorr). The equation below gives the output voltage on the pressure:

V (channel1) = pressure 100

V (channel 2) = 10 × pressure

Where V(channel 1) is the voltage between pins 5 and 6, V(channel 2) is the voltage between pins 7 and 6, and the pressure is indicated in Torr.

HPM-2002-OBEVacuum Gauge

Page 11 of 39

Image 11
Contents Teledyne Hastings Manual Print History Table of Contents Troubleshooting HPM-2002-OBE Analog Output 4-20mA Module FeaturesModel 2002 Sensors HPM-2002-OBE Analog Output ModuleHPM-2002-OBE DeviceNet Module SpecificationsHPM-2002-OBE RS232/485 Output Module Quick Start Transducer InstallationOBE Module Installation Receiving InspectionPin Volt 20 mA RS232/485 Initial OperationHigh and Low Set Point Modes Analog OutputOverall Functional Description Cal Mode Run ModeAnalog Output GAS ModeGas Selection Table HPM-2002-OBEVacuum Gauge Analog Output 4-20mA Cal Mode Analog Output 4-20mA HPM-2002-OBEVacuum Gauge RS232/485 With Display Command SyntaxParamater Modification Commands Interrogation CommandsReset / Restore Commands Calibration Adjustment CommandsDevice Status Default RS232/485 SpecificationsModular Connector Pinout RS-485 Modular Connector Pinout R2-232 RJ-11 connector on left sideRun Mode DeviceNet description DeviceNet TMCalibration of the HPM-2002-OBE Zero Coefficient AdjustmentMidrange Coefficient Adjustment Atmosphere Coefficient AdjustmentTheory of Operation Vo = SPV+V1 Piezoresistive SensorHPM-2002-OBEVacuum Gauge Er = σεTh4-Ta4As Pirani SensorEt = Es + Er + Eg Es = K ∆T Am/LHPM-2002-OBEVacuum Gauge Dual Sensor Operation Eg = Kg ∆T As/∆xEg = arLt273/Th1/2Th-TaAgP HPM-2002-OBEVacuum Gauge RS485 Address Advanced Setup GuideData Bits/Parity Bits/Stop Bits Turnaround DelayRestore Eeprom Baud RateFrequently Asked Questions Why does my display read, Err1? Restore CalibrationRestore Factory Defaults Software ResetWhat are Err3, Err5, Err6 and Err7? Why does my display read, U.U?Why does my display read, U .? Why does my display read, . U?Non-Warranty Repair Policy Warranty Repair PolicyDiagrams and Drawings HPM-2002-OBEVacuum Gauge HPM-2002-OBEVacuum Gauge