Teledyne HPM-2002-OBE instruction manual Interrogation Commands, Paramater Modification Commands

Page 18

The characters within wavy brackets { } contain choices for the appropriate command.

The characters within the symbols < > are the common abbreviations for the one digit ASCII control codes which they represent, (e.g. <CR> represents carriage return).

When entering more than one command in the same data string, they must be separated by a comma.

All command strings must be followed by the terminator character (carriage return <CR>, also known as ENTER).

When a lower case character is present in an example it represents an option.

Character

Description

Valid Inputs:

a

RS-485 Address (hexadecimal 0-9, A-F)

01 - DF

m

Most Significant Digit Of Mantissa

1 - 9

d

Decimal Digit

0 - 9

e

Exponent

0 - 5

u

Unit Of Pressure

T, M, or P

,

Command Separator (comma)

N/A

<CR>

Command Terminator (carriage return)

N/A

3.3.3.Interrogation Commands

Command Description

Format

Sample Response

Transmit Averaged Pressure

P<CR> Pa: 1.23456e+0 Torr<CR>

Transmit Pirani Pressure

R<CR>Pr: 1.98765e-3 Torr<CR>

Transmit Piezo Pressure

Z<CR>Pz: 7.65432e+2 Torr<CR>

Transmit RS-485 Address

A<CR>Multidrop Address: 01<CR>

Transmit Decimation Ratio

D<CR>

Decimation Ratio: 255<CR>

Transmit Selected Gas #

G<CR>

Gas#: 0<CR>

Transmit High Setpoint

H<CR>

Hi: 1.00000e+1 Torr<CR>

Transmit Low Setpoint

L<CR>Lo: 1.00000e-2 Torr<CR>

Transmit Device Status

S<CR> 00044<CR>

Transmit Turnaround Delay

T<CR>

Comm Delay: 6<CR>

Transmit Selected Units

U<CR>

Torr<CR>

Transmit Software Version #

V<CR>Hastings Instruments-OBE 2002

 

 

Version 1.4 - (7-21-00) <CR>

3.3.4.Paramater Modification Commands

Command Description

Format

Valid Range:

 

Modify High Setpoint

H={m.dd}E{+e}<CR> 1.00000e-9 to 9.99999e+9

Modify Low Setpoint

L={m.dd}E{-e}<CR>

1.00000e-9 to 9.99999e+9

Modify Selected Gas #

G={d}<CR>

0 to 4

(Decimal)

Modify Selected Units

U={u}<CR>

T, or M, or P

 

Modify Decimation Ratio

D={dddd}<CR>

63 to 7936

(Decimal)

Modify RS-485 Address

*{aa}A={aa}<CR>

1 to DF (Hexadecimal)

Modify Turnaround Delay

*{aa}T={dd}<CR>

0 to 255(Decimal)

Notes:

HPM-2002-OBEVacuum Gauge

Page 18 of 39

Image 18
Contents Teledyne Hastings Manual Print History Table of Contents Troubleshooting HPM-2002-OBE Analog Output Module FeaturesModel 2002 Sensors HPM-2002-OBE Analog Output 4-20mA ModuleSpecifications HPM-2002-OBE RS232/485 Output ModuleHPM-2002-OBE DeviceNet Module Receiving Inspection Transducer InstallationOBE Module Installation Quick StartInitial Operation Pin Volt 20 mA RS232/485Analog Output Overall Functional DescriptionHigh and Low Set Point Modes Run Mode Cal ModeGAS Mode Gas Selection TableAnalog Output HPM-2002-OBEVacuum Gauge Analog Output 4-20mA Cal Mode Analog Output 4-20mA HPM-2002-OBEVacuum Gauge Command Syntax RS232/485 With DisplayInterrogation Commands Paramater Modification CommandsCalibration Adjustment Commands Reset / Restore CommandsDefault RS232/485 Specifications Device StatusModular Connector Pinout R2-232 RJ-11 connector on left side Modular Connector Pinout RS-485Run Mode DeviceNet TM DeviceNet descriptionAtmosphere Coefficient Adjustment Zero Coefficient AdjustmentMidrange Coefficient Adjustment Calibration of the HPM-2002-OBETheory of Operation Piezoresistive Sensor Vo = SPV+V1HPM-2002-OBEVacuum Gauge Es = K ∆T Am/L Pirani SensorEt = Es + Er + Eg Er = σεTh4-Ta4AsHPM-2002-OBEVacuum Gauge Eg = Kg ∆T As/∆x Eg = arLt273/Th1/2Th-TaAgPDual Sensor Operation HPM-2002-OBEVacuum Gauge Advanced Setup Guide RS485 AddressBaud Rate Turnaround DelayRestore Eeprom Data Bits/Parity Bits/Stop BitsSoftware Reset Restore CalibrationRestore Factory Defaults Frequently Asked Questions Why does my display read, Err1?Why does my display read, . U? Why does my display read, U.U?Why does my display read, U .? What are Err3, Err5, Err6 and Err7?Warranty Repair Policy Non-Warranty Repair PolicyDiagrams and Drawings HPM-2002-OBEVacuum Gauge HPM-2002-OBEVacuum Gauge