Teledyne HPM-2002-OBE What are Err3, Err5, Err6 and Err7?, Why does my display read, U.U?

Page 35

communications program (e.g. WindowsÔ Hyperterminal) which can be used to communicate with RS232 versions of the HPM-2002-OBE. The Transmit Device Status (“S<CR>”) command will retrieve the HPM-2002-OBE’s status register. A successful read of the status register will clear the “Err4” condition.

What are Err3, Err5, Err6 and Err7?

These are simultaneous error conditions, i.e. Err3 = Err1+Err2.

Why does my display read, “U.U”?

“U.U” stands for unconnected piezo/unconnected Pirani. This message indicates that the voltage readings from the sensors are outside any usable range. Most often this occurs when there is no sensor tube connected to the HPM-2002-OBE. This error will also occur if there is a problem with the user’s DC power supply. Verify that the voltage supplied to the HPM-2002-OBE is within the specified range using a voltmeter at pins 3 and 4 on the power cable. After power verification and with the voltage still applied, hot plug the cable into the HPM-2002-OBE. If the “U.U” still appears, most likely the sensor tube (HPM-2002s) needs to be replaced.

Why does my display read, “U .”?

“U.” stands for unconnected piezo. This message indicates that the voltage reading from the piezoresistive sensor is outside any usable range. While this message can mean that the sensor tube (HPM-2002s) will need to be replaced, it can also indicate that the piezo zero calibration parameter, which is stored in the sensor tube’s EEPROM, has become corrupted or lost. (See next FAQ)

Why does my display temporarily show, “U .” as I pump the system down?

“U .” stands for unconnected piezo. This message may flash near the dual sensor crossover pressure range if the piezo zero calibration parameter has been corrupted or lost.

The RESTORE CALIBRATION command will download the calibration parameters from the HPM-2002- OBE to the sensor tube. See the advanced setup guide (Section 5.1) for more information. If the error persists at the crossover point, then a calibration of the Piezo sensor should be performed. The Piezo Zero is updated when the Pirani pressure is less than 125 mTorr. In order to update and save the Piezo Zero parameter:

Pump out to 100 mTorr or less

Using SELECT, step to the CAL function

Rotate ADJUST one step in either direction and then back to its original position (CAL light starts flashing)

Using SELECT, exit the CAL function and return to the Run mode

Changing the Piezo Zero will shift the atmospheric reading by the same amount. therefor the atmospheric readong should also be checked and re-adjusted if necessary.

Why does my display read, “. U”?

This message indicates that the voltage readings from the thin film Pirani sensor are outside any usable range. This usually indicates that the sensor is defective and will need to be replaced.

HPM-2002-OBEVacuum Gauge

Page 35 of 39

Image 35
Contents Teledyne Hastings Manual Print History Table of Contents Troubleshooting HPM-2002-OBE Analog Output 4-20mA Module FeaturesModel 2002 Sensors HPM-2002-OBE Analog Output ModuleHPM-2002-OBE DeviceNet Module SpecificationsHPM-2002-OBE RS232/485 Output Module Quick Start Transducer InstallationOBE Module Installation Receiving InspectionPin Volt 20 mA RS232/485 Initial OperationHigh and Low Set Point Modes Analog OutputOverall Functional Description Cal Mode Run ModeAnalog Output GAS ModeGas Selection Table HPM-2002-OBEVacuum Gauge Analog Output 4-20mA Cal Mode Analog Output 4-20mA HPM-2002-OBEVacuum Gauge RS232/485 With Display Command SyntaxParamater Modification Commands Interrogation CommandsReset / Restore Commands Calibration Adjustment CommandsDevice Status Default RS232/485 SpecificationsModular Connector Pinout RS-485 Modular Connector Pinout R2-232 RJ-11 connector on left sideRun Mode DeviceNet description DeviceNet TMCalibration of the HPM-2002-OBE Zero Coefficient AdjustmentMidrange Coefficient Adjustment Atmosphere Coefficient AdjustmentTheory of Operation Vo = SPV+V1 Piezoresistive SensorHPM-2002-OBEVacuum Gauge Er = σεTh4-Ta4As Pirani SensorEt = Es + Er + Eg Es = K ∆T Am/LHPM-2002-OBEVacuum Gauge Dual Sensor Operation Eg = Kg ∆T As/∆xEg = arLt273/Th1/2Th-TaAgP HPM-2002-OBEVacuum Gauge RS485 Address Advanced Setup GuideData Bits/Parity Bits/Stop Bits Turnaround DelayRestore Eeprom Baud RateFrequently Asked Questions Why does my display read, Err1? Restore CalibrationRestore Factory Defaults Software ResetWhat are Err3, Err5, Err6 and Err7? Why does my display read, U.U?Why does my display read, U .? Why does my display read, . U?Non-Warranty Repair Policy Warranty Repair PolicyDiagrams and Drawings HPM-2002-OBEVacuum Gauge HPM-2002-OBEVacuum Gauge