Teledyne HPM-2002-OBE Analog Output, Overall Functional Description, High and Low Set Point Modes

Page 9

3. Operating Information

This section contains information on operating the HPM-2002-OBE in any of its various configurations. Refer to the appropriate section for information on the mode in use.

3.1. Analog Output (0-10V)

3.1.1.Overall Functional Description

The status bar gives information about the condition of the HPM-2002-OBE. High and Low indicate whether the set points are activated. ERROR indicates if there was a problem downloading the EEPROM. Torr and mTorr indicate what pressure regime the gauge is measuring.

The mode bar indicates which mode has been selected. The HPM-2002-OBE has five modes which the user may enter. This is similar to the six modes which can be entered in the bench top/panel mounted HPM- 2002, there is no interlock feature; when the user changes the setting within a mode, any adjustments that have been made will be permanent once the mode is exited. To exit without making changes permanent, the user must turn the power off.

3.1.2.High and Low Set Point Modes

The HPM-2002-OBE provides TTL outputs for process control. These signals are available on the 15-pin connector (see previous table in Section 2.4).

To view the High set point, place the HPM-2002-OBE in the High mode by pressing the SELECT button until the High mode light is illuminated. The display then shows the set point selected. During normal operation the alarm light will illuminate and the TTL output (pin # 1) will go high (+5V) if the pressure exceeds the set point.

Similarly, to view the Low set point, place the HPM-2002-OBE in the Low mode by pressing the SELECT button until the Low mode light is illuminated. The display then shows the set point selected. During normal operation the alarm light will illuminate and the TTL output (pin # 2) will go high (+5V) if the pressure becomes less than the set point.

To adjust a setpoint, place the HPM-2002-OBE in either setpoint mode (High or Low). Next, use the ADJUST rotary encoder until the desired setpoint is displayed. Finally, place the HPM-2002-OBE back in the Run mode. The new setpoints are now stored in the HPM-2002-OBE’s memory.

HPM-2002-OBEVacuum Gauge

Page 9 of 39

Image 9
Contents Teledyne Hastings Manual Print History Table of Contents Troubleshooting Model 2002 Sensors FeaturesHPM-2002-OBE Analog Output Module HPM-2002-OBE Analog Output 4-20mA ModuleSpecifications HPM-2002-OBE RS232/485 Output ModuleHPM-2002-OBE DeviceNet Module OBE Module Installation Transducer InstallationReceiving Inspection Quick StartPin Volt 20 mA RS232/485 Initial OperationAnalog Output Overall Functional DescriptionHigh and Low Set Point Modes Cal Mode Run ModeGAS Mode Gas Selection TableAnalog Output HPM-2002-OBEVacuum Gauge Analog Output 4-20mA Cal Mode Analog Output 4-20mA HPM-2002-OBEVacuum Gauge RS232/485 With Display Command SyntaxParamater Modification Commands Interrogation CommandsReset / Restore Commands Calibration Adjustment CommandsDevice Status Default RS232/485 SpecificationsModular Connector Pinout RS-485 Modular Connector Pinout R2-232 RJ-11 connector on left sideRun Mode DeviceNet description DeviceNet TMMidrange Coefficient Adjustment Zero Coefficient AdjustmentAtmosphere Coefficient Adjustment Calibration of the HPM-2002-OBETheory of Operation Vo = SPV+V1 Piezoresistive SensorHPM-2002-OBEVacuum Gauge Et = Es + Er + Eg Pirani SensorEs = K ∆T Am/L Er = σεTh4-Ta4AsHPM-2002-OBEVacuum Gauge Eg = Kg ∆T As/∆x Eg = arLt273/Th1/2Th-TaAgPDual Sensor Operation HPM-2002-OBEVacuum Gauge RS485 Address Advanced Setup GuideRestore Eeprom Turnaround DelayBaud Rate Data Bits/Parity Bits/Stop BitsRestore Factory Defaults Restore CalibrationSoftware Reset Frequently Asked Questions Why does my display read, Err1?Why does my display read, U .? Why does my display read, U.U?Why does my display read, . U? What are Err3, Err5, Err6 and Err7?Non-Warranty Repair Policy Warranty Repair PolicyDiagrams and Drawings HPM-2002-OBEVacuum Gauge HPM-2002-OBEVacuum Gauge